Literature DB >> 2523244

Fracture of a balloon on a wire device during coronary angioplasty.

M van den Brand1, P de Feyter, P Serruys, F Zijlstra, E Bos.   

Abstract

In a 61-year-old patient with unstable angina an attempt was made to dilate a severe stenosis in a tortuous obtuse marginal branch. The initial attempt with conventional equipment was not successful; although the wire could be advanced distal to the stenosis, a 2.0 balloon did not cross the stenosis. A second attempt with a balloon on a wire device resulted in fracture of this catheter, with the distal 2.8-cm-long fragment looped in the left coronary artery. Immediate bypass surgery was performed and the broken fragment was easily removed from the left coronary ostium. The patient made an uneventful recovery.

Entities:  

Mesh:

Year:  1989        PMID: 2523244     DOI: 10.1002/ccd.1810160410

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569


  1 in total

1.  Surgical removal of entrapped guidewire fragment following direct circumflex artery angioplasty and stenting.

Authors:  Wojciech Domaradzki; Krzysztof Sanetra; Bartosz Skwarna; Marek Król; Małgorzata Kuczera; Marek Cisowski
Journal:  Kardiochir Torakochirurgia Pol       Date:  2015-09-28
  1 in total