| Literature DB >> 25226973 |
Wei-chen Lee1, Ching-Hang Huang1, Shan-Chen Chung1, Ching-chih Wei1.
Abstract
The aim of this research was to develop an efficient and accurate method to fabricate a dental implant surgical guide. The surgical guide is adapted from the patient's vacuum-formed clear template with the use of a plate with three ceramic balls, a six-axis drilling machine and its fixture. The plate, with the ceramic balls used as radiographic markers, is glued to the template, and the patient bites this template during a CT scan. Then, the surgeon can plan the locations and orientations of the implants on the CT-based model in the dental planning software. The drilling information is exported directly to the computer-controlled drilling machine for subsequent drilling on the template to complete the surgical guide. This method allows the surgical guide to be made without any measurements, which reduces the fabrication time, but increases the drilling accuracy. The preliminary results show that the average location error was 0.31 ± 0.17 mm and the average orientation error was 0.53 ± 0.24°, which can be considered accurate in comparison with the results reported in the literature.Entities:
Keywords: Dental implant surgical guide; drilling machine; fixture; radiographic marker; vacuum-formed template
Mesh:
Substances:
Year: 2014 PMID: 25226973 DOI: 10.3233/BME-141086
Source DB: PubMed Journal: Biomed Mater Eng ISSN: 0959-2989 Impact factor: 1.300