OBJECTIVE: In the surgery of calcaneal fracture, whether the sustentaculum tali screw should always be placed is widely controversial. The aim of this study was to explore the necessity and function of the sustentaculum tali screw placement for the treatment of Sanders type II calcaneal fracture.Entities:
Keywords: Calcaneal fracture; Sustentaculum tali; Screws; Finite element analysis; Biomechanics
Year: 2014 PMID: 25225534 PMCID: PMC4163240 DOI: 10.12669/pjms.305.5301
Source DB: PubMed Journal: Pak J Med Sci ISSN: 1681-715X Impact factor: 1.088
The elastic constants of the different materials
|
|
|
|
|---|---|---|
| Cortical bone | 7300 | 0.3 |
| Cancellous bone | 100 | 0.3 |
| Titanium plate and screws | 200000 | 0.28 |
| Fracture line | 5 | 0.4 |
The load and nodes in each surface in the neutral position of the calcaneus
|
|
|
|
|---|---|---|
| Contact surface between the Achilles tendon and ground | 1620 | Constraint |
| Calcaneocuboid articular surface | 950 | Constraint |
| Middle subtalar articular surface | 1348 | 200 |
| Posterior subtalar articular surface | 1682 | 420 |
| Calcaneal tuberosity | 1563 | -300 |
Fig.1The displacement of the fracture line. A. The minimum displacement of the fracture line in the model with sustentaculum tali screw was 9.0946×10-7 mm and the maximum displacement was 0.23986 mm. B. The minimum displacement in the model without sustentaculum tali screw was 1.2241×10-6 mm and the maximum displacement was 0.24938mm
Fig.2The stress distribution in the calcaneus. A. The maximum stress in the model with sustentaculum tali screw was 78.582 MPa, which was mainly concentrated on the cortical bone around the sustentaculum tali. B. The maximum stress in the model without sustentaculum tali screw was 95.427 MPa, which was also mainly concentrated on the cortical bone around the sustentaculum tali
Fig.3The stress distribution in the implants. A. The maximum stress in the implants with sustentaculum tali screw was 284.64 MPa, which was mainly concentrated in the middle part of the plate. B. The maximum stress in the implants without sustentaculum tali screw was 186.95 MPa, which was mainly concentrated in the posterior subtalar surface around the tail of the cancellous screws