Literature DB >> 25220268

Reference characterisation of sound speed and attenuation of the IEC agar-based tissue-mimicking material up to a frequency of 60 MHz.

Srinath Rajagopal1, Neelaksh Sadhoo2, Bajram Zeqiri2.   

Abstract

To support the development of clinical applications of high-frequency ultrasound, appropriate tissue-mimicking materials (TMMs) are required whose acoustic properties have been measured using validated techniques. This paper describes the characterisation of the sound speed (phase velocity) and attenuation coefficient of the International Electrotechnical Commission (IEC) agar-based TMM over the frequency range 1 to 60 MHz. Measurements implemented a broadband through-transmission substitution immersion technique over two overlapping frequency ranges, with co-axially aligned 50 MHz centre-frequency transducers employed for characterisation above 15 MHz. In keeping with usual practice employed within the technical literature, thin acoustic windows (membranes) made of 12-μm-thick Mylar protected the TMM from water damage. Various important sources of uncertainty that could compromise measurement accuracy have been identified and evaluated through a combination of experimental studies and modelling. These include TMM sample thickness, measured both manually and acoustically, and the influence of interfacial losses that, even for thin protective membranes, are significant at the frequencies of interest. In agreement with previous reports, the attenuation coefficient of the IEC TMM exhibited non-linear frequency dependence, particularly above 20 MHz, yielding a value of 0.93 ± 0.04 dB cm(-1) MHz(-1) at 60 MHz, derived at 21 ± 0.5°C. For the first time, phase velocity, measured with an estimated uncertainty of ±3.1 m s(-1), has been found to be dispersive over this extended frequency range, increasing from 1541 m s(-1) at 1 MHz to 1547 m s(-1) at 60 MHz. This work will help standardise acoustic property measurements, and establishes a reference measurement capability for TMMs underpinning clinical applications at elevated frequencies. Crown
Copyright © 2015. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Attenuation coefficient; Frequency dependence; Speed of sound; Tissue-mimicking material; Ultrasound

Mesh:

Substances:

Year:  2014        PMID: 25220268     DOI: 10.1016/j.ultrasmedbio.2014.04.018

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  7 in total

1.  A Comparison of the Sensitivity of Contrast-Specific Imaging Modes on Clinical and Preclinical Ultrasound Scanners.

Authors:  Carmel M Moran; Charles Arthur; Emilio Quaia
Journal:  Tomography       Date:  2022-09-15

2.  Evaluation of ultrasonic scattering in agar-based phantoms using 3D printed scattering molds.

Authors:  Antria Filippou; Christakis Damianou
Journal:  J Ultrasound       Date:  2022-01-08

3.  Broadband characterization of plastic and high intensity therapeutic ultrasound phantoms using time delay spectrometry-With validation using Kramers-Kronig relations.

Authors:  Subha Maruvada; Yunbo Liu; Paul Gammell; Keith Wear
Journal:  J Acoust Soc Am       Date:  2018-06       Impact factor: 1.840

4.  In vitro characterisation of ultrasound-induced heating effects in the mother and fetus: A clinical perspective.

Authors:  Stephanie F Smith; Piero Miloro; Richard Axell; Gail Ter Haar; Christoph Lees
Journal:  Ultrasound       Date:  2020-09-14

5.  Assessment of Spectral Doppler for an Array-Based Preclinical Ultrasound Scanner Using a Rotating Phantom.

Authors:  David A Kenwright; Tom Anderson; Carmel M Moran; Peter R Hoskins
Journal:  Ultrasound Med Biol       Date:  2015-05-06       Impact factor: 2.998

6.  acoustic assessment of a konjac–carrageenan tissue-mimicking material aT 5–60 MHZ.

Authors:  David A Kenwright; Neelaksh Sadhoo; Srinath Rajagopal; Tom Anderson; Carmel M Moran; Patrick W Hadoke; Gillian A Gray; Bajram Zeqiri; Peter R Hoskins
Journal:  Ultrasound Med Biol       Date:  2014-12       Impact factor: 2.998

7.  Validation of hybrid angular spectrum acoustic and thermal modelling in phantoms.

Authors:  Sara L Johnson; Douglas A Christensen; Christopher R Dillon; Allison Payne
Journal:  Int J Hyperthermia       Date:  2018-10-15       Impact factor: 3.914

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.