| Literature DB >> 25214289 |
G Fiard1, M Peneau2, B Soulimane2.
Abstract
The sliding-clip renorraphy technique using Hem-o-lok(®) clips for parenchymal closure in partial nephrectomy is largely used in laparoscopic, robotic and sometimes open surgery. We present here a case of massive migration of the clips after open partial nephrectomy. We discuss the imaging characteristics of those clips: not visible on plain X-ray, low density on computed tomography, before reviewing previously published cases.Keywords: Cancer du rein; Complication; Foreign body migration; Imagerie; Imaging; Kidney cancer; Migration de corps étranger; Néphrectomie partielle; Partial nephrectomy; Ureteroscopy; Urétéroscopie
Mesh:
Year: 2014 PMID: 25214289 DOI: 10.1016/j.purol.2014.05.001
Source DB: PubMed Journal: Prog Urol ISSN: 1166-7087 Impact factor: 0.915