The impact of CyberKnife's prescription isodose percentage on intracranial target planning.

Sung-Woo Lee1, Sunyoung Jang, Anil P Pyakuryal, Kenneth Chang, Terence T Sio.   

Abstract

Entities:  

Mesh:

Year:  2014        PMID: 25207584      PMCID: PMC5711081          DOI: 10.1120/jacmp.v15i5.5081

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


× No keyword cloud information.
To the Editor: Recently, a detailed comparative study regarding intracranial Gamma Knife (GK) vs. CyberKnife (CK) intracranial dosimetry has been published by your Journal. In a group of 15 patients with 26 brain metastases, we showed that CK produced more homogeneous and conformal plans, while the GK plans had sharper peripheral dose falloff in most cases. In the CK plans, by convention, the applied range of prescription isodose percentage (PIP) was 77%–92%, with a median value of 86%. Intrigued by the results, we hypothesized that lowering the PIP in CK planning would improve peripheral dose falloff, without compromising the excellent dosimetric conformality which had previously been achieved. Secondarily, it was expected that, as PIP decreased, the stereotactic radiosurgical (SRS) plans would become less homogeneous as maximum dose within the target increased. We thank you for the opportunity to share with you the additional results that were generated from this investigation. Parts of the methods and materials have been previously described. We compared the relative dosimetric merit of various prescription isodose levels in CK's MultiPlan (Accuray Inc, Sunnyvale, CA). The same 15‐patient series was used for dosimetric planning. For each tumor, the PIP was varied at three levels averaging approximately 50, 65, and 85% (CK50, CK65 and CK85; Table 1). The homogeneity (HI) and gradient (GI) indices, modified conformity index (mCI, the ratio of the prescription isodose volume to the tumor volume receiving at least the prescription dose), and an MPS‐defined quantity called “new CI” (nCI, the ratio of mCI to target coverage, also the inverse of van't Riet's Conformation Number) were computed. For peripheral dose falloff, GI50 was calculated as the ratio of the volume enclosed by the isodose at 50% of the prescription dose level to the volume enclosed by the original prescription isodose. GI25, GI40, GI60, and GI80 were calculated in a similar manner. Statistical analyses were performed using analysis of variance (ANOVA) and nonparametric Kruskal‐Wallis tests.
Table 1

Tested patient's tumor location, volume, and PIP

Final PIP Chosen (%) CK65
Patient No. Tumor Location Volume of Tumors (cm 3 ) CK50 CK65 CK85
1Left Frontal1.2506488
Right Cerebellar1.4506878
Vermis1.7506679
Left Occipital2.1506881
2Right Frontal1.2506987
3Right Parietal0.3516886
4Right Cerebellar7.6506590
5Right Frontal5.9526586
6Left Temporal0.1506988
7Left Parietal2.9536889
Parietal Fossa4.8486588
8Left Anterior Frontal19.9506691
Left Post Frontal5.1527092
9Left Cerebellar7.0506583
Left Frontal1.2506577
10Vermis5.8506685
Right Cerebellar11.5496584
11Right IAC0.5526891
Middle Cerebellar0.7506988
12Right Frontal2.1506981
Left Temporal0.4526883
Right Frontal4.6546582
13Right Post Frontal3.6506583
14Vermis14.5496682
Brainstem1.5506589
15Right Frontal8.9486579
Median (Range)2.5 (0.1–19.9)50 (48–54)66 (64–70)86 (77–92)

; ; .

We found that the mean tumor volume was 4.4 cm3; a median dose of 18 Gy was prescribed. For CK50, CK65, and CK85 series, the coverage was maintained at 96%–100% in all cases. Optimized plans in each scenario across various PIPs were computed, and dosimetric constraints of critical organ structures were all met. Minimum, average, maximum doses, HI, nCI, GI25, GI50, and MU were reported (Table 2). Comparing across the CK50, CK65, and CK85 series, the median mCIs were: 1.48, 1.36, and 1.52, , respectively. The remaining gradient indices were: GI40 (5.8, 6.9, and 7.6, ); for GI60 (2.8, 3.4, and 3.8), GI80 (1.6, 1.9, and 2.2), and GI90 (1.3, 1.4, and 1.6), in all cases. In our study, as expected, the selection of a PIP had a statistically significant impact on HI, mean, and maximum doses. By both mCI and nCI, CK65 produced the most conformal plans which nearly reached statistical significance. Importantly, CK50 had significantly sharpest dose falloff at all gradient index levels, with the exception of GI25. However, dosimetric plans prescribing to CK50 required significantly longer treatment times by MU estimation.
Table 2

Summary of CyberKnife's dosimetric parameters (mean values) as varied by prescription isodose levels

CK50 CK65 CK85 F(H)‐ratio p‐value
Minimum Dose (Gy)17.617.317.60.10.92
Mean Dose (Gy)27.122.920.530.8 <0.00001 a
Maximum Dose (Gy)36.927.922.1102.5 <0.00001 a
Homogeneity Index (HI)1.991.501.181568.1 <0.00001 a
New Conformity Index (nCI)1.501.381.552.60.09
Gradient Index 25 (GI25)15.616.716.30.50.80
Gradient Index 50 (GI50)3.84.75.222.3 <0.0001 a
Monitor Unit (MU)20084139521202512.40.0021 a

Statistically significant (two‐sided ).

.

In current clinical practices, deciding a prescription isodose level in CK varies by individual plan and preference, and clearly no consensus exists. The CK is a relatively new modality for SRS, which is stereotactically capable for extracranial indications, as well. For GK and linac‐based intracranial SRS, it is common to prescribe to 40%–60% and 80%–95% of PIP, respectively. It is then customarily believed that the PIPs of CK should fall between those of the GK and linac‐based SRS plans, as CK shares features of both. For CK dosimetric planning, some authors used median PIPs of about 79%, while others believed values near 73% may be more appropriate. Inoue et al. used 60% PIP for a three‐fractioned course of 27 Gy. Our current data evaluated a wider range of PIPs compared to these published studies. Additionally, some researchers tried to compare different radiosurgery modalities by interpreting the equivalent uniform biologic effective dose of normal brain tissue for GK, CK, and linac‐based systems, with largely equivalent results. Petti et al. suggested dosimetric guidance for GK users to optimize CI to reduce normal brain toxicity derived by empirical formula. According to their study, CI was an important dosimetric parameter to consider when tumor size exceeded 1 cm3, which is equally important in CK planning. Tested patient's tumor location, volume, and PIP ; ; . Summary of CyberKnife's dosimetric parameters (mean values) as varied by prescription isodose levels Statistically significant (two‐sided ). . In terms of dose falloff, as Sahgal et al. and Ma reported, the choice of PIP was an important factor to consider in plan quality optimization. Particularly, the authors noted that, in patients with multiple metastases, the choice of PIP became critical as SRS treatment extended to increased number of target sites. In our study, we also observed similar peripheral dose falloff variations as a result of different PIPs. Specific to CK, the multiplan system (MPS) has three optimization functions to emulate GK‐like “isocentric” forward planning, in additional to IMRT‐like multi‐isocentric “conformal” and “sequential” inverse planning algorithms; the optional selection of such a dosimetric function, especially in relation to varying PIP, should be the topic of a future study. We conclude that, for CK dosimetric planning in focused irradiation of intracranial targets, the choosing of a PIP had a statistically significant impact on HI, minimum, mean, and maximum doses. By both mCI and nCI criteria, CK65 produced the most conformal plans which nearly reached statistical significance. Although CK50 plans required longer treatment times as estimated by MU, their peripheral dose falloff was sharpest at all gradient index levels, except for GI25. PIP is an important variable to consider in CK planning; tradeoff among homogeneity, treatment time, and dose gradient falloff must be carefully balanced in individualized stereotactic radiosurgical plans.
  8 in total

1.  Dependence of normal brain integral dose and normal tissue complication probability on the prescription isodose values for gamma-knife radiosurgery.

Authors:  L Ma
Journal:  Phys Med Biol       Date:  2001-11       Impact factor: 3.609

2.  Prescription dose guideline based on physical criterion for multiple metastatic brain tumors treated with stereotactic radiosurgery.

Authors:  Arjun Sahgal; Igor J Barani; Josef Novotny; Beibei Zhang; Paula Petti; David A Larson; Lijun Ma
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-05-14       Impact factor: 7.038

3.  Equivalence in dose fall-off for isocentric and nonisocentric intracranial treatment modalities and its impact on dose fractionation schemes.

Authors:  Lijun Ma; Arjun Sahgal; Martina Descovich; Young-Bin Cho; Cynthia Chuang; Kim Huang; Normand J Laperriere; Dennis C Shrieve; David A Larson
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-01       Impact factor: 7.038

4.  A theoretical investigation of optimal target-dose conformity in gamma knife radiosurgery.

Authors:  Paula L Petti; Sandeep Kunwar; David A Larson
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

5.  CyberKnife enhanced conventionally fractionated chemoradiation for high grade glioma in close proximity to critical structures.

Authors:  Eric Oermann; Brian T Collins; Kelly T Erickson; Xia Yu; Sue Lei; Simeng Suy; Heather N Hanscom; Joy Kim; Hyeon U Park; Andrew Eldabh; Christopher Kalhorn; Kevin McGrail; Deepa Subramaniam; Walter C Jean; Sean P Collins
Journal:  J Hematol Oncol       Date:  2010-06-09       Impact factor: 17.388

6.  Stereotactic radiosurgery of the postoperative resection cavity for brain metastases.

Authors:  Scott G Soltys; John R Adler; John D Lipani; Paul S Jackson; Clara Y H Choi; Putipun Puataweepong; Scarlett White; Iris C Gibbs; Steven D Chang
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-19       Impact factor: 7.038

7.  Comparing gamma knife and cyberknife in patients with brain metastases.

Authors:  Terence T Sio; Sunyoung Jang; Sung-Woo Lee; Bruce Curran; Anil P Pyakuryal; Edward S Sternick
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

8.  Three-fraction CyberKnife radiotherapy for brain metastases in critical areas: referring to the risk evaluating radiation necrosis and the surrounding brain volumes circumscribed with a single dose equivalence of 14 Gy (V14).

Authors:  Hiroshi K Inoue; Ken-Ichi Seto; Akihiko Nozaki; Kota Torikai; Yoshiyuki Suzuki; Jun-Ichi Saitoh; Shin-Ei Noda; Takashi Nakano
Journal:  J Radiat Res       Date:  2013-02-11       Impact factor: 2.724

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.