Ns Senthil Kumar1, Baharudin Omar2, Leonard H Joseph1, Nor Hamdan1, Ohnmar Htwe3, Nursalbiyah Hamidun1. 1. Physiotherapy Program, School of Rehabilitation Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, Malaysia. 2. Department of Biomedical Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, Malaysia. 3. Department of Orthopedic and Traumatology, Faculty of Medicine, Universiti Kebangsaan Malaysia Medical Centre, Malaysia.
Abstract
[Purpose] The aim of the present study was to investigate the accuracy of a digital weight scale relative to the Wii in limb loading measurement during static standing. [Methods] This was a cross-sectional study conducted at a public university teaching hospital. The sample consisted of 24 participants (12 with osteoarthritis and 12 healthy) recruited through convenient sampling. Limb loading measurements were obtained using a digital weight scale and the Nintendo Wii in static standing with three trials under an eyes-open condition. The limb load asymmetry was computed as the symmetry index. [Results] The accuracy of measurement with the digital weight scale relative to the Nintendo Wii was analyzed using the receiver operating characteristic (ROC) curve and Kolmogorov-Smirnov test (K-S test). The area under the ROC curve was found to be 0.67. Logistic regression confirmed the validity of digital weight scale relative to the Nintendo Wii. The D statistics value from the K-S test was found to be 0.16, which confirmed that there was no significant difference in measurement between the equipment. [Conclusion] The digital weight scale is an accurate tool for measuring limb load asymmetry. The low price, easy availability, and maneuverability make it a good potential tool in clinical settings for measuring limb load asymmetry.
[Purpose] The aim of the present study was to investigate the accuracy of a digital weight scale relative to the Wii in limb loading measurement during static standing. [Methods] This was a cross-sectional study conducted at a public university teaching hospital. The sample consisted of 24 participants (12 with osteoarthritis and 12 healthy) recruited through convenient sampling. Limb loading measurements were obtained using a digital weight scale and the Nintendo Wii in static standing with three trials under an eyes-open condition. The limb load asymmetry was computed as the symmetry index. [Results] The accuracy of measurement with the digital weight scale relative to the Nintendo Wii was analyzed using the receiver operating characteristic (ROC) curve and Kolmogorov-Smirnov test (K-S test). The area under the ROC curve was found to be 0.67. Logistic regression confirmed the validity of digital weight scale relative to the Nintendo Wii. The D statistics value from the K-S test was found to be 0.16, which confirmed that there was no significant difference in measurement between the equipment. [Conclusion] The digital weight scale is an accurate tool for measuring limb load asymmetry. The low price, easy availability, and maneuverability make it a good potential tool in clinical settings for measuring limb load asymmetry.
Entities:
Keywords:
Accuracy; Digital weight scale; Nintendo Wii Fit
Limb load asymmetry (LLA) is common in clinical conditions related to the lower limbs such
as degenerative joint disease, soft-tissue injuries, joint replacements, amputations, and
stroke1, 2). LLA is defined as an uneven distribution of body weight on one
lower extremity when compared with the other extremity2). LLA may occur due to pain, reduced muscle strength, and/or impaired
joint mobility1). Eventually, LLA may cause
degeneration of the joint and osteopenia1).
Hence, LLA may be an important clinical variable for investigation of impairments and the
need for further rehabilitation.A variety of equipment is available to measure LLA, including biofeedback systems,
ambulatory devices with sensors, the Tekscan MatScan system, and force platforms3). These pieces of equipment provide better
accuracy and reliable LLA measurements. However, they are rarely available in practice due
to their high costs and are confined to laboratory settings. In addition, these pieces of
equipment also require ample space and specialized training to handle them in clinical
settings.The aforementioned limitations of these pieces of equipment highlight the need for a
simple, inexpensive device to measure LLA in daily practice. Recent studies have utilized a
visual simulation gaming tool, such as the Nintendo Wii (Wii), to measure LLA3, 4).
The Wii provides reliable and valid LLA measurements when compared with the gold standard
force platform (ICC = 0.77–0.89)5).
Nevertheless, the Wii requires a television for data capture, and it is less mobile. Recent
literature has proposed measurement of LLA using 2 pieces of digital weight scale (DWS)
equipment4). The DWS is inexpensive,
portable, and easy to utilize in the clinical setting3, 4). However, the accuracy of LLA
measurement obtained from the DWS is uncertain. Accuracy refers to how close a value is to
the accepted value5). It is indicated as
the measurement closeness to truth. In the present study, accuracy refers to the closeness
of agreement between the limb loading measurement values from the DWS to the measurement
values from the Wii. Hence, this study focuses on the accuracy of the DWS by comparison with
the Wii.
SUBJECTS AND METHODS
This experimental study was conducted in the human movements and health research unit of a
public university teaching hospital. The study samples comprised 12 osteoarthritis Kellgren-
Lawrence grade 2 and 3 and 12 healthy participants. Participants were recruited if they
could understand commands and stand independently and were without any neurological
disorders. Ethical approval was obtained from the university research committee. Prior to
data collection, informed consent was obtained from all participants. LLA was measured using
two Beurer DWSs (BEU-GS27-0007, Beurer, Ulm, Germany), each with a precision of 0.01 kg. The
maximum measurement capability of each DWS was 150 kg. Prior to data collection, the
accuracy of the two DWSs was tested with dead weights to ensure zero errors2). The two DWSs were placed on flat, hard
flooring at a distance equivalent to the participant’s shoulder width6). Participants were instructed to stand in a relaxed posture
while facing forward with their bare feet centrally placed on each DWS7). The readings were noted after standing on the scales for 10
seconds. Three trials of measurement were noted and averaged for data analysis.The Nintendo Wii Fit system (Nintendo of America Inc., Redmond, WA, USA) consists of a
balance board, console, remote, and television for data acquisition. The Wii console was
used to synchronize the data from the balance board to a television. The measurement
procedure used for the DWS was repeated to collect data from the Wii. The participants were
given 5 minutes of rest between DWS and Wii measurements. The sequences of equipments were
randomized. One researcher collected the data throughout the study. The loading measurements
from both lower limbs were computed for LLA analysis using the symmetry index (SI)8). The SI is derived from a mathematical
formula that provides LLA values in percentages.Accuracy was analyzed using the Statistical Package for the Social Sciences (version 20). A
receiver operating characteristic (ROC) curve was utilized to compare the measurement data
obtained from the Wii and DWS. Two- sample Kolmogorov-Smirnov (K-S) tests were utilized in
conjunction with the ROC, as this test is sensitive to differences in measurement values,
for both the location and shape of the empirical cumulative distribution functions of the
DWS and Wii.
RESULTS
The study group consisted of 24 participants, 12 osteoarthritis (OA) participants (8
female, 4 male) and 12 healthy participants (8 female, 4 male). The mean age of the
osteoarthritisparticipants was 61.25± 6.68 yrs, mean body mass index (BMI) was
27.26±3.36 kg/m2 and mean Visual Analogue Scale (VAS) score was 4.83±1.59. The
mean age of the healthy participants was 61.83 ± 7.86 yrs, and their mean BMI was 26.71 ±
4.76 kg/m2. Based on the VAS scores, none of the healthy participants reported
pain in the lower limbs.The ROC curve was used to test the accuracy of the DWS relative to the Wii. The area under
the curve was 0.67 with a 95% confidence interval (p=0.06). The upper bound and lower bound
values were 0.50 and 0.82 respectively. The standard error of measurement was 0.08% of
change between the symmetry indexes of the DWS and Wii. The K-S test results are shown in
Table 1.
Table 1.
Comparison of SI values from the DWS and Wii using the two-sample
Kolmogorov-Smirnov (K-S) test
Mean ± SD
95% CI
High
Low
SI (DWS)
4.16 ± 2.7
3.02 − 5.30
0.10
1.33
SI (Wii)
4.16 ± 2.9
2.93 − 5.39
0.11
2.38
DISCUSSION
In practice, existence of asymmetry in lower limb loading among healthy populations is a
controversial issue. Measurement of limb loading during while standing is crucial in various
orthopedic and neurological conditions1, 8, 9).
However, the current measurement tools are not readily available in clinical practice.
Hence, the present study proposed the use of two DWSs for measurement of LLA and tested the
accuracy relative to the Nintendo Wii.The results of the present study demonstrate the measurement accuracy of the DWS relative
to the Wii. The area under the ROC curve was found to be 0.65, which is between 0.5 to 1.0,
indicating a best fit between the two pieces of equipment. In addition, the logistic
regression line indicates that the results are not by chance. Hence, the DWS and Wii produce
similar LLA measurements during a standing task. According to the K-S test, the maximum
difference between the cumulative distributions of the SI obtained from the DWS and Wii is
given by a D value of 0.16, which is greater than the critical value (p = 0.86). This proves
that the symmetry index scores obtained from the DWS and Wii are comparable.The study findings could be ascribed to the technical configurations of the two pieces of
equipment, as the DWS and Nintendo Wii balance board uses strain gauge sensors to sense load
shifts5). They could also be attributed
to the calibration and zero error checking conducted for the two DWSs prior to data
collection. The two DWSs used in this study had a precision of 0.01 kg, which indicates that
the closeness of agreement between independent measurements might have reduced the chance of
uncertainty. Also, this study used three trails, and the measurements were averaged for
computation, which might have reduced the random error, as supported by the central limit
theorem10). Finally, the standard
procedure and instructions given to the participants during the measurements might have been
responsible for the similarities in measurements between the devices. The DWS could fill the
gap with regard to tools for measurement of LLA based on its advantages in terms of
portability, ease of operation, instant readings, and proven accuracy. The findings of the
present study could indicate opportunities for clinicians and researchers to decide on the
usefulness of the DWS as a clinical and research tool for measurement of LLA in the field of
rehabilitation.