Bacterial Composition at the Implant-Abutment Connection under Loading in vivo.

Georgios E Romanos1, Mary Therese Biltucci2, Alexis Kokaras3, Bruce J Paster3,4.   

Abstract

PURPOSE: Platform-switched implants have been demonstrated to prevent bone loss after loading. The present study evaluated bacterial composition of sites from implant-abutment connections of immediately loaded implants, which were placed in the anterior mandible. Ten patients participated in this study.
MATERIALS AND METHODS: A and B implant systems with two different prosthetic connections (Morse tapered vs internal polygonal butt-joint connections, respectively) were placed and loaded for 2 years. The abutments were removed (AB sample) after careful decontamination. Bacterial sampling of the abutments, inner part of the implants (before/visit 1 and after rinsing with chlorhexidine [CHX]/visit 2), and after new abutment connection and loading for 1 additional month, a new sampling (visit 3) was taken to compare the bacteria composition in association with the two connections. Bacterial profiles of samples were determined by using the human oral microbe identification microarray.
RESULTS: A total of 240 samples were analyzed taken at different time intervals. Nonparametric statistical analysis (Wilcoxon Rank sum) with uncorrected alpha (p < .05) and after corrections (Benjamini-Hochberg) found no statistical significance between the two connections. No significant changes in the overall microbial profiles were detected at the different time intervals. However, there were trends toward presence of periodontitis-associated species at the B implants in all samples (AB, visit 1, even after CHX irrigation) and after decontamination, abutment replacement, and 1-month loading period.
CONCLUSIONS: CHX irrigation does not seem to have any effect on decontamination of connections. As shown previously, there is significantly more bone loss around B implants compared with A implants. Although there was no statistical difference in the microbial profiles, there was indeed a trend for the presence of typical periodontal pathogens associated with the internal polygonal butt-joint connection. A possible scenario is that this connection tends to harbor the pathogens that may be involved in subsequent bone loss.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  bacteria, immediate loading; implant-abutment interface; platform switching

Mesh:

Substances:

Year:  2014        PMID: 25195711     DOI: 10.1111/cid.12270

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  7 in total

1.  In vitro analysis of the microbiological sealing of tapered implants after mechanical cycling.

Authors:  Deceles Cristina Costa Alves; Paulo Sérgio Perri de Carvalho; Carlos Nelson Elias; Eduardo Vedovatto; Elizabeth Ferreira Martinez
Journal:  Clin Oral Investig       Date:  2016-02-20       Impact factor: 3.573

Review 2.  Environmental, Microbiological, and Immunological Features of Bacterial Biofilms Associated with Implanted Medical Devices.

Authors:  Marina Caldara; Cristina Belgiovine; Eleonora Secchi; Roberto Rusconi
Journal:  Clin Microbiol Rev       Date:  2022-01-19       Impact factor: 50.129

3.  Volumetric Changes in Morse Taper Connections After Implant Placement in Dense Bone. In-Vitro Study.

Authors:  Georgios E Romanos; Rafael Arcesio Delgado-Ruiz; Ana I Nicolas-Silvente
Journal:  Materials (Basel)       Date:  2020-05-16       Impact factor: 3.623

4.  Peri-Implant Bone Resorption during Healing Abutment Placement: The Effect of a 0.20% Chlorhexidine Gel vs. Placebo-A Randomized Double Blind Controlled Human Study.

Authors:  Bruna Sinjari; Gianmaria D'Addazio; Ilaria De Tullio; Tonino Traini; Sergio Caputi
Journal:  Biomed Res Int       Date:  2018-10-16       Impact factor: 3.411

5.  Sustained co-delivery of ibuprofen and basic fibroblast growth factor by thermosensitive nanoparticle hydrogel as early local treatment of peri-implantitis.

Authors:  Wenlei Chen; Min Zhi; Zujian Feng; Pengfei Gao; Yuan Yuan; Congcong Zhang; Yonglan Wang; Anjie Dong
Journal:  Int J Nanomedicine       Date:  2019-02-21

6.  The effect of dynamic loading on bacterial microleakage of the dental implant fixture-abutment interface: An in vitro study.

Authors:  Mehrak Amjadi; Azadeh Khajavi; Navid Youssefi; Negar Barjini; Zahra Kazemi
Journal:  J Indian Prosthodont Soc       Date:  2021 Oct-Dec

7.  In Vitro Bacterial Leakage at the Implant-Abutment Connection of Two Dental Implant Systems with Internal Connection.

Authors:  Azadeh Khajavi; Shabnam Mohseni; Amir Peymani; Mehrak Amjadi
Journal:  Front Dent       Date:  2020-12-08
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.