Ying Ying Wu1, Anton Plakseychuk2, Kenji Shimada3.
Abstract
Current external fixators for distraction osteogenesis (DO) are unable to correct all types of deformities in the lower limb and are difficult to use because of the lack of a pre-surgical planning system. We propose a DO system that consists of a surgical planner and a new, easy-to-setup unilateral fixator that not only corrects all lower limb deformity, but also generates the contralateral/predefined bone shape. Conventionally, bulky constructs with six or more joints (six degrees of freedom, 6DOF) are needed to correct a 3D deformity. By applying the axis-angle representation, we can achieve that with a compact construct with only two joints (2DOF). The proposed system makes use of computer-aided design software and computational methods to plan and simulate the planned procedure. Results of our stress analysis suggest that the stiffness of our proposed fixator is comparable to that of the Orthofix unilateral external fixator. We tested the surgical system on a model of an adult deformed tibia and the resulting bone trajectory deviates from the target bone trajectory by 1.8mm, which is below our defined threshold error of 2mm. We also extracted the transformation matrix that defines the deformity from the bone model and simulated the planned procedure.Entities:
Keywords: Axial fixator; Axis-angle representation; Bone shape regeneration; Distraction osteogenesis; External fixator; Surgical planner
Mesh:
Year: 2014 PMID: 25194468 DOI: 10.1016/j.medengphy.2014.08.006