Measurement of time delays in gated radiotherapy for realistic respiratory motions.

Brige P Chugh1, Sarah Quirk1, Leigh Conroy1, Wendy L Smith1.   

Abstract

PURPOSE: Gated radiotherapy is used to reduce internal motion margins, escalate target dose, and limit normal tissue dose; however, its temporal accuracy is limited. Beam-on and beam-off time delays can lead to treatment inefficiencies and/or geographic misses; therefore, AAPM Task Group 142 recommends verifying the temporal accuracy of gating systems. Many groups use sinusoidal phantom motion for this, under the tacit assumption that use of sinusoidal motion for determining time delays produces negligible error. The authors test this assumption by measuring gating time delays for several realistic motion shapes with increasing degrees of irregularity.
METHODS: Time delays were measured on a linear accelerator with a real-time position management system (Varian TrueBeam with RPM system version 1.7.5) for seven motion shapes: regular sinusoidal; regular realistic-shape; large (40%) and small (10%) variations in amplitude; large (40%) variations in period; small (10%) variations in both amplitude and period; and baseline drift (30%). Film streaks of radiation exposure were generated for each motion shape using a programmable motion phantom. Beam-on and beam-off time delays were determined from the difference between the expected and observed streak length.
RESULTS: For the system investigated, all sine, regular realistic-shape, and slightly irregular amplitude variation motions had beam-off and beam-on time delays within the AAPM recommended limit of less than 100 ms. In phase-based gating, even small variations in period resulted in some time delays greater than 100 ms. Considerable time delays over 1 s were observed with highly irregular motion.
CONCLUSIONS: Sinusoidal motion shapes can be considered a reasonable approximation to the more complex and slightly irregular shapes of realistic motion. When using phase-based gating with predictive filters even small variations in period can result in time delays over 100 ms. Clinical use of these systems for patients with highly irregular patterns of motion is not advised due to large beam-on and beam-off time delays.

Entities:  

Mesh:

Year:  2014        PMID: 25186377     DOI: 10.1118/1.4890604

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Technical Note: High temporal resolution characterization of gating response time.

Authors:  Rodney D Wiersma; Bradley P McCabe; Andrew H Belcher; Patrick J Jensen; Brett Smith; Bulent Aydogan
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

2.  Real-time liver tracking algorithm based on LSTM and SVR networks for use in surface-guided radiation therapy.

Authors:  Guangyu Wang; Zhibin Li; Guangjun Li; Guyu Dai; Qing Xiao; Long Bai; Yisong He; Yaxin Liu; Sen Bai
Journal:  Radiat Oncol       Date:  2021-01-14       Impact factor: 3.481

3.  Accuracy of real-time respiratory motion tracking and time delay of gating radiotherapy based on optical surface imaging technique.

Authors:  Li Chen; Sen Bai; Guangjun Li; Zhibin Li; Qing Xiao; Long Bai; Changhu Li; Lixun Xian; Zhenyao Hu; Guyu Dai; Guangyu Wang
Journal:  Radiat Oncol       Date:  2020-07-10       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.