An ultra-rigid close-stacked piezo motor for harsh condition scanning probe microscopy.

Ying Guo1, Yubin Hou, Qingyou Lu.   

Abstract

We designed and produced a nearly closest packed stack motor with only one tiny gap of 0.15 mm in the middle of the stack. A low-voltage method of controlling the motor is introduced for the first time. Besides, the test results of the motor and the corresponding scanning tunneling microscope are also presented. To our surprise, it turns out to be so rigid that even running two oil pumps and one ultrasonic cleaner within 1 m range from a STM directly driven by this new motor cannot cause the STM to produce any visible difference in its the atomic resolution quality. This is a leap in building a true harsh condition atomic resolution SPM.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  STM in noise environment; closest packed stack motor; high rigidity; low-voltage control method; minimum mechanical loop

Year:  2014        PMID: 25183527     DOI: 10.1002/sca.21149