Literature DB >> 25181288

Saturation of the right-leg drive amplifier in low-voltage ECG monitors.

Daniel K Freeman, Ronald D Gatzke, Georgios Mallas, Yu Chen, Chris J Brouse.   

Abstract

Electrocardiogram (ECG) monitoring is a critical tool in patient care, but its utility is often balanced with frustration from clinicians who are constantly distracted by false alarms. This has motivated the need to readdress the major factors that contribute to ECG noise with the goal of reducing false alarms. In this study, we describe a previously unreported phenomenon in which ECG noise can result from an unintended interaction between two systems: 1) the dc lead-off circuitry that is used to detect whether electrodes fall off the patient; and 2) the right-leg drive (RLD) system that is responsible for reducing ac common-mode noise that couples into the body. Using a circuit model to study this interaction, we found that in the presence of a dc lead-off system, even moderate increases in the right-leg skin-electrode resistance can cause the RLD amplifier to saturate. Such saturation can produce ECG noise because the RLD amplifier will no longer be capable of attenuating ac common-mode noise on the body. RLD saturation is particularly a problem for modern ECG monitors that use low-voltage supply levels. For example, for a 12-lead ECG and a 2 V power supply, saturation will occur when the right-leg electrode resistance reaches only 2 MΩ. We discuss several design solutions that can be used in low-voltage monitors to avoid RLD saturation.

Entities:  

Mesh:

Year:  2014        PMID: 25181288     DOI: 10.1109/TBME.2014.2351611

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  1 in total

1.  An Electronic System for the Contactless Reading of ECG Signals.

Authors:  Francesca Romana Parente; Marco Santonico; Alessandro Zompanti; Mario Benassai; Giuseppe Ferri; Arnaldo D'Amico; Giorgio Pennazza
Journal:  Sensors (Basel)       Date:  2017-10-28       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.