Device closure of secundum atrial septal defect (ASD) is the treatment of choice when anatomy is favourable. Amplatzer device has remained the gold standard for closure of ASD. Cobra deformity is a well-reported problem with devices. Recently, Tulip deformity has been reported in a single case. We report a series of cases where we noted Tulip deformity along with inability to retract the device in the sheath in Cera Lifetech devices. This resulted in prolongation of procedure, excessive fluoroscopic exposure and additional interventional procedures not usually anticipated in ASD device closure. We believe that the problem is due to the stiffness of the device resulting in its inability to be retracted into the sheath. We also report a unique way of retrieving the device.Entities: