| Literature DB >> 25173320 |
T S Tripathi1, M Bala1, K Asokan1.
Abstract
We report on an experimental setup for the simultaneous measurement of the thermoelectric power (TEP) of two samples in the temperature range from 77 K to 500 K using optimum electronic instruments. The setup consists of two rectangular copper bars in a bridge arrangement for sample mounting, two surface mount (SM) chip resistors for creating alternate temperature gradient, and a type E thermocouple in differential geometry for gradient temperature (ΔT) measurement across the samples. In addition, a diode arrangement has been made for the alternate heating of SM resistors using only one DC current source. The measurement accuracy of ΔT increases with the differential thermocouple arrangement. For the calibration of the setup, measurements of TEP on a high purity (99.99%) platinum wire and type K thermocouple wires Chromel and Alumel have been performed from 77 K to 500 K with respect to copper lead wires. Additionally, this setup can be utilized to calibrate an unknown sample against a sample of known absolute TEP.Entities:
Year: 2014 PMID: 25173320 DOI: 10.1063/1.4891631
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523