Literature DB >> 25167877

Comparison of analog and digital transceiver systems for MR imaging.

Seitaro Hashimoto1, Katsumi Kose, Tomoyuki Haishi.   

Abstract

We critically evaluated analog and digital transceivers for magnetic resonance (MR) imaging systems under identical experimental conditions to identify and compare their advantages and disadvantages. MR imaging experiments were performed using a 4.74-tesla vertical-bore superconducting magnet and a high sensitivity gradient coil probe. We acquired 3-dimensional spin echo images of a kumquat with and without using a gain-stepping scan technique to extend the dynamic range of the receiver systems. The acquired MR images clearly demonstrated nearly identical image quality for both transceiver systems, but DC and ghosting artifacts were obtained for the analog transceiver system. We therefore concluded that digital transceivers have several advantages over the analog transceivers.

Mesh:

Year:  2014        PMID: 25167877     DOI: 10.2463/mrms.2013-0114

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  1 in total

1.  Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil.

Authors:  Nao Kodama; Katsumi Kose
Journal:  Magn Reson Med Sci       Date:  2016-03-21       Impact factor: 2.471