| Literature DB >> 25167877 |
Seitaro Hashimoto1, Katsumi Kose, Tomoyuki Haishi.
Abstract
We critically evaluated analog and digital transceivers for magnetic resonance (MR) imaging systems under identical experimental conditions to identify and compare their advantages and disadvantages. MR imaging experiments were performed using a 4.74-tesla vertical-bore superconducting magnet and a high sensitivity gradient coil probe. We acquired 3-dimensional spin echo images of a kumquat with and without using a gain-stepping scan technique to extend the dynamic range of the receiver systems. The acquired MR images clearly demonstrated nearly identical image quality for both transceiver systems, but DC and ghosting artifacts were obtained for the analog transceiver system. We therefore concluded that digital transceivers have several advantages over the analog transceivers.Mesh:
Year: 2014 PMID: 25167877 DOI: 10.2463/mrms.2013-0114
Source DB: PubMed Journal: Magn Reson Med Sci ISSN: 1347-3182 Impact factor: 2.471