Comparison of different approaches to atrioventricular junction ablation and pacemaker implantation in patients with atrial fibrillation.

Nisha Arenja1, Sven Knecht, Beat Schaer, Tobias Reichlin, Nikola Pavlovic, Stefan Osswald, Christian Sticherling, Michael Kühne.   

Abstract

BACKGROUND: To compare the feasibility and efficiency of atrioventricular junction (AVJ) ablation and device implantation in patients with drug-refractory atrial fibrillation using three different approaches.
METHODS: Sixty-nine patients (57% male; age 72 ± 10; ejection fraction 45 ± 15%) undergoing device implantation and AVJ ablation were retrospectively studied at a tertiary referral center. In 20 patients (29%) AVJ ablation was performed via the femoral vein immediately following device implantation (group 1), whereas 33 patients (48%) underwent a staged procedure with AVJ ablation via the femoral vein >3 weeks after device implantation (group 2). In a third group of 16 patients (23%), AVJ ablation was performed during device implantation through the pocket using the same axillary vein access site (group 3). The main outcome measures were: procedure time, fluoroscopy time, laboratory occupancy time, and success rate.
RESULTS: There was a significant difference in procedure time (118 ± 45 minutes. in group 1, 133 ± 32 minutes in group 2, and 87 ± 26 minutes in group 3, P < 0.001) and the laboratory occupancy time (175 ± 48 minutes in group 1, 200 ± 32 minutes in group 2, and 121 ± 27 minutes in group 3, P < 0.001). There was no difference in fluoroscopy time (group 1: 20 ± 15 minutes, group 2: 27 ± 22 minutes, and group 3: 24 ± 9 minutes P = 0.4). The procedure was successfully completed in all patients, but cross-over to a femoral approach was required in one patient in group 3.
CONCLUSION: The alternative approach of AVJ ablation during permanent pacemaker implantation from the same axillary vein access site is feasible and more efficient compared to the femoral approach. ©2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  alternative approach; atrial fibrillation; atrioventricular junction ablation; axillary vein access site; pacemaker implantation

Mesh:

Year:  2014        PMID: 25160503     DOI: 10.1111/pace.12481

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  SANS FLUORO Too Big to Fail.

Authors:  Robert L Percell; Matthew E Johnson; Raghuveer Dendi
Journal:  J Atr Fibrillation       Date:  2021-02-28

2.  Strategy of pacemaker and electrode replacement for superior vena cava stenosis.

Authors:  Zhi-Peng Zheng; Wei Zhang; Zhi-Huan Zeng; Jian-Yi Zheng
Journal:  J Int Med Res       Date:  2021-01       Impact factor: 1.671

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.