Literature DB >> 25158119

Computed tomographic analysis of left ventricular volumes and function after implantation of the Parachute® endoventricular partitioning device.

Prahlad G Menon1, Daniel Ludwig, Joan Lacomis, David Schwartzman, Catalin Toma.   

Abstract

BACKGROUND: Preliminary clinical experience with a percutaneous endoventricular partitioning device (Parachute®, CardioKinetix Inc., Menlo Park, CA, USA) suggests that it ameliorates global LV dysfunction and heart failure symptoms in selected patients who have suffered previous anterior myocardial infarction. Less is known of its effect on regional LV function.
OBJECTIVE: To gain insight into device effect on regional LV function by analysis of cardiac computed tomographic (CT) images obtained before and after device implantation.
METHODS: Comparative analysis of pre and 6 months post-implantation contrast-enhanced CT images from 6 subjects enrolled in the phase 1 Parachute clinical trials, including regional LV volume and systolic excursion, as well as device motion.
RESULTS: After implantation, a significant reduction in volume of the "dynamic" LV compartment (that which was not excluded by the device) was accompanied by a significant reduction in dykinetic motion and a trend toward an improved ejection fraction. Penetration of contrast into the excluded compartment was still present at 6 months, however the apical motion was significantly less dyskinetic in 3 subjects and unchanged in the other 3. Overall device surface motion was inward in systole, a significant improvement relative to the overall dyskinetic LV apex pre-implantation. Device motion was spatially heterogeneous, which appeared to be dependent on the motility of the myocardium that anchored its individual splines.
CONCLUSIONS: Our data suggest that the Parachute device acts as a functional impediment to flow and stretch, effectively depressurizing the apical segment.
© 2014, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2014        PMID: 25158119     DOI: 10.1111/joic.12151

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  1 in total

1.  cMRI-BED: A novel informatics framework for cardiac MRI biomarker extraction and discovery applied to pediatric cardiomyopathy classification.

Authors:  Vanathi Gopalakrishnan; Prahlad G Menon; Shobhit Madan
Journal:  Biomed Eng Online       Date:  2015-08-13       Impact factor: 2.819

  1 in total