Literature DB >> 25153415

Use of quality indicators to compare point-of-care testing errors in a neonatal unit and errors in a STAT central laboratory.

Miguel Cantero, Maximino Redondo, Eva Martín, Gonzalo Callejón, María Luisa Hortas.   

Abstract

BACKGROUND: Point-of-care testing (POCT), like other laboratory tests, can be affected by errors throughout the total testing process. To evaluate quality error rates, the use of quality indicators (QIs) is recommended; however, little information is available on the quality error rate associated with POCT. The objective of this study was to investigate quality error rates related to POCT and compare them with central laboratory (CL) testing.
METHODS: We studied standardized QIs for POCT in comparison to CL testing. We compared error rates related to requests, collection, and handling of samples and results from external quality assessment program (EQAP) and internal quality control (IQC).
RESULTS: The highest difference between POCT and CL testing was observed for QI related to patient identification, 45.3% vs. 0.02% (p<0.001). Regarding specimen collection and handling, the QI related to samples without results was also higher in POCT than in CL testing, 15.8% vs. 3.3% (p<0.001). For the QI related to insufficient sample volume, we obtained 2.9% vs. 0.9% (p=0.27). Unlike QIs for the preanalytical phase, QIs for the analytical phase had better results in POCT than CL testing. We obtained 8.3% vs. 16.6% (p=0.13) for QI related to unacceptable results in EQAP and 0.8% vs. 22.5% (p<0.001) for QI related to unacceptable results in IQC.
CONCLUSIONS: Our results show that the preanalytical phase remains the main problem in POCT like in CL testing and that monitoring of quality indicators is a very valuable tool in reducing errors in POCT.

Entities:  

Mesh:

Year:  2015        PMID: 25153415     DOI: 10.1515/cclm-2013-1053

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Real-world use of key performance indicators for point-of-Care Testing network accredited by ISO 22870.

Authors:  Paloma Oliver; Pilar Fernandez-Calle; Roberto Mora; Jorge Diaz-Garzon; Daniel Prieto; Marta Manzano; Inmaculada Dominguez; Antonio Buño
Journal:  Pract Lab Med       Date:  2020-11-06

2.  Identifying sources of error and selecting quality indicators for point of care testing.

Authors:  Miranda Brun; Anna K Füzéry; Bailey Henschke; Kallie Rozak; Allison A Venner
Journal:  Pract Lab Med       Date:  2021-03-21

Review 3.  Practical challenges related to point of care testing.

Authors:  Julie L V Shaw
Journal:  Pract Lab Med       Date:  2015-12-09

4.  Glucose point-of-care meter operators competency: An assessment checklist.

Authors:  Cindy Tang Friesner; Julien Meyer; Pria Nippak
Journal:  Pract Lab Med       Date:  2020-03-05
  4 in total