Anterior knee symptoms after S-ROM hinge implantation.

David J Deehan, Rajkumar Gangadharan, Ajay Malviya, Alasdair Sutherland, James P Holland.   

Abstract

PURPOSE: To evaluate the performance of a canal filling hinge device for complex knee arthroplasty.
METHODS: Thirty-seven (4 primary hinge implantation and 33 revision cases) patients who had undergone arthroplasty with the S-ROM third generation hinge device for a combination of massive bone loss or ligamentous insufficiency were prospectively examined with a minimum of 5-year follow-up. Median age at surgery was 72 years (range: 43 to 87 years). Principal indications included aseptic loosening or massive osteolysis (24 cases), infection (8 cases) and periprosthetic fracture (4 cases). All patients exhibited either grade 2 (N = 12) or grade 3 (N = 25) AORI bone loss or a grade 3 medial ligament deficiency.
RESULTS: One patient experienced implant failure (71 months), and one patient suffered late deep infection (36 months). Mean WOMAC score improved from 27 to 62. Four patients required patellar resurfacing for persistent pain. The 5-year survivorship was 86%.
CONCLUSIONS: While the S-ROM device may offer satisfactory medium term outcome for complex end stage knee disease, we report a high rate of debilitating anterior knee symptoms.

Entities:  

Mesh:

Year:  2014        PMID: 25150346

Source DB:  PubMed          Journal:  Bull Hosp Jt Dis (2013)        ISSN: 2328-4633


  2 in total

Review 1.  Total knee arthroplasty using hinge joints: Indications and results.

Authors:  E Carlos Rodríguez-Merchán
Journal:  EFORT Open Rev       Date:  2019-04-25

2.  High ten-year implant survivorship and low patellofemoral complication rate for S-ROM rotating-hinge implants in revision total knee arthroplasty : a single-centre study.

Authors:  Hosam E Matar; Benjamin V Bloch; Peter J James
Journal:  Bone Jt Open       Date:  2022-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.