Literature DB >> 25130948

Svelte integrated delivery system performance examined through diagnostic catheter delivery: the SPEED registry.

Ahmed A Khattab1, Freek Nijhoff, Joachim Schofer, Jacques Berland, Bernhard Meier, Fabian Nietlispach, Pierfrancesco Agostoni, Steffen Brucks, Pieter Stella.   

Abstract

AIMS: The multi-center SPEED registry evaluated the procedural success and in-hospital clinical outcomes of direct stenting with the Svelte 'all-in-one' coronary stent Integrated Delivery System (IDS) through diagnostic catheters to identify the clinical indications for which this approach is appropriately suited. METHODS &
RESULTS: Forty-eight (48) patients with 54 lesions of lengths ≤20 mm and RVD 2.5-3.5 mm were targeted for direct stenting through diagnostic catheters (4-6F) via radial or femoral approach. Procedural characteristics early in an investigator's experience (28 lesions) were compared with outcomes following experience (26 lesions). Procedure, device and strategy success were realized in 54 (100%), 50 (93%) and 46 (85%) lesions, respectively, with strategy success significantly related to RVD (P = 0.05), lesion location (P = 0.01), and diagnostic catheter size (P = 0.05). Significant improvement in crossing and intervention time and trends toward improvement in device and strategy success, reductions in procedure and radiation time and contrast use were observed.
CONCLUSIONS: Direct stenting through diagnostic catheters via radial or femoral approach using the Svelte IDS is feasible and associated with good in-hospital outcomes. This approach offers the attractive option of assessing lesions via diagnostic catheter and, depending upon vessel anatomy and lesion morphology, continuing with ad-hoc interventional treatment using the same diagnostic catheter. Improvements in strategy success and procedural efficiencies, based on operator experience, facilitate catheter downsizing and reduce intervention time, ancillary product use and overall procedure costs.
© 2014 Wiley Periodicals, Inc. © 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  coronary artery disease; diagnostic cardiac catheterization; direct stenting; percutaneous coronary intervention

Mesh:

Year:  2014        PMID: 25130948     DOI: 10.1002/ccd.25621

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

1.  Current State of Bioabsorbable Polymer-Coated Drug-Eluting Stents.

Authors:  Abhilash Akinapelli; Jack P Chen; Kristine Roy; Joseph Donnelly; Keith Dawkins; Barbara Huibregtse; Dongming Hou
Journal:  Curr Cardiol Rev       Date:  2017
  1 in total