| Literature DB >> 25121654 |
H Hachisu, M Fujieda, S Nagano, T Gotoh, A Nogami, T Ido, St Falke, N Huntemann, C Grebing, B Lipphardt, Ch Lisdat, D Piester.
Abstract
We have demonstrated a direct frequency comparison between two ⁸⁷Sr lattice clocks operated in intercontinentally separated laboratories in real time. Two-way satellite time and frequency transfer technique, based on the carrier-phase, was employed for a direct comparison, with a baseline of 9000 km between Japan and Germany. A frequency comparison was achieved for 83,640 s, resulting in a fractional difference of (1.1±1.6)×10⁻¹⁵, where the statistical part is the largest contributor to the uncertainty. This measurement directly confirms the agreement of the two optical frequency standards on an intercontinental scale.Year: 2014 PMID: 25121654 DOI: 10.1364/OL.39.004072
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776