Liqiang Liu1, Changrui Xing2, Huijuan Yan3, Hua Kuang4, Chuanlai Xu2.
Abstract
A monoclonal antibody for microcystin-leucine-arginine (MC-LR) was produced by cell fusion. The immunogen was synthesized in two steps. First, ovalbumin/ bovine serum albumin was conjugated with 6-acetylthiohexanoic acid using a carbodiimide EDC (1-ethyl-3-[3-dimethylaminopropyl]carbodiimide hydrochloride)/ NHS (N-hydroxysulfosuccinimide) reaction. After dialysis, the protein was reacted with MC-LR based on a free radical reaction under basic solution conditions. The protein conjugate was used for immunization based on low volume. The antibodies were identified by indirect competitive (ic)ELISA and were subjected to tap water and lake water analysis. The concentration causing 50% inhibition of binding of MC-LR (IC50) by the competitive indirect ELISA was 0.27 ng/mL. Cross-reactivity to the MC-RR, MC-YR and MC-WR was good. The tap water and lake water matrices had no effect on the detection limit. The analytical recovery of MC-LR in the water samples in the icELISA was 94%-110%. Based on this antibody, an immunochromatographic biosensor was developed with a cut-off value of 1 ng/mL, which could satisfy the requirement of the World Health Organization for MC-LR detection in drinking water. This biosensor could be therefore be used as a fast screening tool in the field detection of MC-LR.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25120158 PMCID: PMC4179011 DOI: 10.3390/s140814672
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.Standard curve for measurement of MC-LR.
IC50 and cross-reactivities.
| MC-LR(Microcystin-Leucine-Arginine) | 0.27 | 100% |
| [D-Asp3] MC-RR([D-Asp3] Microcystin-Arginine-Arginine) | 0.18 | 150% |
| MC-YR(Microcystin-Tyrosine-Arginine) | 0.22 | 123% |
| MC-RR(Microcystin-Arginine-Arginine) | 0.28 | 96% |
| MC-WR(Microcystin-Tryptophan-Arginine) | 0.29 | 93% |
| MC-HtyR(Microcystin-Homotyrosine-Arginine) | 0.6 | 45% |
| MC-LY(Microcystin-Leucine-Tyrosine) | >300 | <0.1% |
| MC-LW(Microcystin-Leucine-Tryptophan) | >300 | <0.1% |
| MC-LF(Microcystin-Leucine-Phenylalanine) | >300 | <0.1% |
| MC-LA(Microcystin-Leucine-Alanine) | >300 | <0.1% |
MC-LR recovery values from spiked water sample by ELISA.
| Drinking water | 0.2 | 0.207 | 103 |
| 0.5 | 0.523 | 105 | |
| Tap water | 0.2 | 0.187 | 94 |
| 0.5 | 0.532 | 107 | |
| Lake water | 0.2 | 0.212 | 106 |
| 0.5 | 0.553 | 110 |
Intra- and inter-assay variation for MC-LR detection.
|
|
| |||
|---|---|---|---|---|
| 0.2 | 3 | 3.9 | 5 | 5.2 |
| 0.5 | 3 | 2.1 | 5 | 4.3 |
Intra-assay variation was calculated from 3 replicates on a single day.
Inter-assay variation was calculated from triplicates on 5 different days.
CV: coefficient of variation.
Figure 2.Schematic diagram of the immunochromatography for MC-LR detection.
Figure 3.Typical photo image of detection MC-LR by strip sensor in PBS.
Figure 4.Typical photo image of detection MC-LR by strip sensor in tap water (A) and lake water (B).
Performance Strip Test.
|
| |||
|---|---|---|---|
| 0 | + | + | + |
| 1 | − | − | − |
Negative result. The test line is obviously observed.
Positive result. No test line is observed.