Neonatal and adult ICU ventilators to provide ventilation in neonates, infants, and children: a bench model study.

Laurence Vignaux1, Lise Piquilloud2, Pierre Tourneux3, Philippe Jolliet2, Peter C Rimensberger4.   

Abstract

BACKGROUND: Using a bench test model, we investigated the hypothesis that neonatal and/or adult ventilators equipped with neonatal/pediatric modes currently do not reliably administer pressure support (PS) in neonatal or pediatric patient groups in either the absence or presence of air leaks.
METHODS: PS was evaluated in 4 neonatal and 6 adult ventilators using a bench model to evaluate triggering, pressurization, and cycling in both the absence and presence of leaks. Delivered tidal volumes were also assessed. Three patients were simulated: a preterm infant (resistance 100 cm H2O/L/s, compliance 2 mL/cm H2O, inspiratory time of the patient [TI] 400 ms, inspiratory effort 1 and 2 cm H2O), a full-term infant (resistance 50 cm H2O/L/s, compliance 5 mL/cm H2O, TI 500 ms, inspiratory effort 2 and 4 cm H2O), and a child (resistance 30 cm H2O/L/s, compliance 10 mL/cm H2O, TI 600 ms, inspiratory effort 5 and 10 cm H2O). Two PS levels were tested (10 and 15 cm H2O) with and without leaks and with and without the leak compensation algorithm activated.
RESULTS: Without leaks, only 2 neonatal ventilators and one adult ventilator had trigger delays under a given predefined acceptable limit (1/8 TI). Pressurization showed high variability between ventilators. Most ventilators showed TI in excess high enough to seriously impair patient-ventilator synchronization (> 50% of the TI of the subject). In some ventilators, leaks led to autotriggering and impairment of ventilation performance, but the influence of leaks was generally lower in neonatal ventilators. When a noninvasive ventilation algorithm was available, this was partially corrected. In general, tidal volume was calculated too low by the ventilators in the presence of leaks; the noninvasive ventilation algorithm was able to correct this difference in only 2 adult ventilators.
CONCLUSIONS: No ventilator performed equally well under all tested conditions for all explored parameters. However, neonatal ventilators tended to perform better in the presence of leaks. These findings emphasize the need to improve algorithms for assisted ventilation modes to better deal with situations of high airway resistance, low pulmonary compliance, and the presence of leaks.
Copyright © 2014 by Daedalus Enterprises.

Entities:  

Keywords:  artificial; equipment safety; intensive care units; mechanical ventilators; models; neonatal; pediatrics; respiration; ventilatory support

Mesh:

Year:  2014        PMID: 25118306     DOI: 10.4187/respcare.02540

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  5 in total

1.  A new infant hybrid respiratory simulator: preliminary evaluation based on clinical data.

Authors:  Barbara Stankiewicz; Krzysztof J Pałko; Marek Darowski; Krzysztof Zieliński; Maciej Kozarski
Journal:  Med Biol Eng Comput       Date:  2017-03-25       Impact factor: 2.602

2.  Comparative bench study evaluation of different infant interfaces for non-invasive ventilation.

Authors:  Giorgio Conti; Giorgia Spinazzola; Cesare Gregoretti; Giuliano Ferrone; Andrea Cortegiani; Olimpia Festa; Marco Piastra; Luca Tortorolo; Roberta Costa
Journal:  BMC Pulm Med       Date:  2018-04-07       Impact factor: 3.317

3.  The HEV Ventilator: at the interface between particle physics and biomedical engineering.

Authors:  Jan Buytaert; Paula Collins; Adam Abed Abud; Phil Allport; Antonio Pazos Álvarez; Kazuyoshi Akiba; Oscar Augusto de Aguiar Francisco; Aurelio Bay; Florian Bernard; Sophie Baron; Claudia Bertella; Josef X Brunner; Themis Bowcock; Martine Buytaert-De Jode; Wiktor Byczynski; Ricardo De Carvalho; Victor Coco; Ruth Collins; Nikola Dikic; Nicolas Dousse; Bruce Dowd; Kārlis Dreimanis; Raphael Dumps; Paolo Durante; Walid Fadel; Stephen Farry; Antonio Fernàndez Prieto; Arturo Fernàndez Tèllez; Gordon Flynn; Vinicius Franco Lima; Raymond Frei; Abraham Gallas Torreira; Tonatiuh García Chàvez; Evangelos Gazis; Roberto Guida; Karol Hennessy; Andre Henriques; David Hutchcroft; Stefan Ilic; Artūrs Ivanovs; Aleksandar Jevtic; Emigdio Jimenez Dominguez; Christian Joram; Kacper Kapusniak; Edgar Lemos Cid; Jana Lindner; Rolf Lindner; M Ivàn Martínez Hernàndez; Mirko Meboldt; Marko Milovanovic; Sylvain Mico; Johan Morant; Michel Morel; Georg Männel; Dónal Murray; Irina Nasteva; Niko Neufeld; Igor Neuhold; Francisco Pardo-Sobrino López; Eliseo Pèrez Trigo; Gonzalo Pichel Jallas; Edyta Pilorz; Lise Piquilloud; Xavier Pons; David Reiner; Hector David Règules Medel; Saul Rodríguez Ramírez; Mario Rodíguez Cahuantzi; Carl Roosens; Philipp Rostalski; Freek Sanders; Eric Saucet; Marianne Schmid Daners; Burkhard Schmidt; Patrick Schoettker; Rainer Schwemmer; Heinrich Schindler; Archana Sharma; Derick Sivakumaran; Christophe Sigaud; Vasilios Spitas; Nicola Steffen; Peter Svihra; Guillermo Tejeda Muñoz; Nikolaos Tachatos; Efstratios Tsolakis; Jan van Leemput; Laurence Vignaux; Francois Vasey; Hamish Woonton; Ken Wyllie
Journal:  R Soc Open Sci       Date:  2022-03-16       Impact factor: 2.963

4.  Effect of tidal volume and end tracheal tube leakage on end-tidal CO2 in very low birth weight infants.

Authors:  Daijiro Takahashi; Koko Goto; Kei Goto
Journal:  J Perinatol       Date:  2020-08-04       Impact factor: 2.521

5.  Comparative bench study evaluation of a modified snorkeling mask used during COVID-19 pandemic and standard interfaces for non-invasive ventilation.

Authors:  Giuliano Ferrone; Giorgia Spinazzola; Roberta Costa; Antonio Gullì; Andrea Scapigliati; Massimo Antonelli; Giorgio Conti
Journal:  Pulmonology       Date:  2021-06-08
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.