Craig Clifford1, Scott Berg2, Kevin McCann3, Byron Hutchinson4.
Abstract
The purpose of the present study was to compare the primary bending stiffness characteristics of 5 different ankle arthrodesis fixation techniques: 3 compression screws, an anterior locking plate, a lateral locking plate, an anterior locking plate with a compression screw, and a lateral locking plate with a compression screw. A total of 25 full-scale anatomic models consisting of fourth-generation composite tibiae and tali were tested using an Instron 4505 Universal Testing System. We hypothesized that the use of a compression screw with a locking plate would add considerable stiffness to the fixation construct compared with the use of a locking plate alone. The data have shown that an anterior or lateral plate with a compression screw provides significantly greater stiffness than both a plate and 3 compression screws used individually. No significant difference was seen between the anterior plate with a compression screw and the lateral plate with a compression screw. No significant differences were found among the use of an anterior plate, a lateral plate, or 3 compression screws. We have concluded that when using a locking plate in an anterior or lateral configuration, the addition of a compression screw will considerably increase the primary bending stiffness of ankle arthrodesis.Entities: