Literature DB >> 2511400

Measurement of dose distributions using film in therapeutic electron beams.

A S Shiu1, V A Otte, K R Hogstrom.   

Abstract

The feasibility of using film dosimetry data as the input data for patient treatment planning was evaluated. The central-axis depth dose and the off-axis ratios obtained from film measurements in a solid phantom were compared with those of ion-chamber measurements in water. Two techniques were used to generate isodose distributions. The first technique used only the film data, i.e., the central-axis depth dose and the off-axis ratios used for the reconstruction were determined from the film optical density (corrected for film nonlinearity). In the second technique, the central-axis depth dose measured by an ion chamber in a water phantom was combined with the off-axis ratios measured using film in the "solid water" phantom. The resulting isodose distributions from both techniques were compared with the ion-chamber measurements in water for 7-, 12-, and 18-MeV electrons, and the second technique showed better agreement with the ion-chamber measurements than did the first technique. The differences were within a clinically acceptable range.

Entities:  

Mesh:

Year:  1989        PMID: 2511400     DOI: 10.1118/1.596409

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  An integrated treatment delivery system for CSRS and CSRT and clinical applications.

Authors:  A Shiu; B Parker; J-S Ye; J Lii
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102