Viscoelastic properties of common suture material used for rotator cuff repair and arthroscopic procedures.

Daniel Hurwit1, Gary Fanton1, Mallik Tella1, Anthony Behn1, Kenneth J Hunt2.   

Abstract

PURPOSE: The purpose of this study was to evaluate the viscoelastic properties of 5 suture materials, commonly used in arthroscopic rotator cuff repairs, when subjected to physiological loads.
METHODS: We evaluated 5 commercially available No. 2 sutures undergoing both creep and cyclic testing in both dry air and phosphate-buffered saline solution (PBS) maintained at 37°C. The selected sutures were MagnumWire (ArthroCare, Austin, TX), Ethibond (Ethicon, Somerville, NJ), FiberWire (Arthrex, Naples, FL), Orthocord (DePuy, Warsaw, IN), and Force Fiber (Tornier, Bloomington, MN).
RESULTS: Regarding creep testing, in the PBS test environment, FiberWire showed the greatest stiffness (71.1 ± 2.1 N/mm), the smallest initial extension at the 60-N load (1.10 ± 0.04 mm), and the smallest amount of creep (0.57 ± 0.01 mm). Orthocord showed the smallest amount of relaxed elongation in PBS (0.73 ± 0.11 mm). Regarding cyclic testing, in the PBS testing environment, Ethibond exhibited the smallest dynamic creep (0.28 ± 0.02 mm), FiberWire displayed the smallest peak-to-peak displacement (0.17 ± 0.00 mm), and Orthocord showed the smallest amount of relaxed elongation after cyclic loading (0.63 ± 0.11 mm).
CONCLUSIONS: FiberWire consistently displayed more extreme viscoelastic properties--greater stiffness and less extensibility--than the other suture types studied. Orthocord showed the smallest amount of relaxed elongation in both testing environments. Differences in testing environment affect the behavior of each suture type. Testing in physiologically approximating conditions such as PBS maintained at 37°C is warranted. CLINICAL RELEVANCE: Although many other factors affect the success of rotator cuff repairs, the viscoelastic properties of sutures may be a useful predictor of suture performance.
Copyright © 2014 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25108903     DOI: 10.1016/j.arthro.2014.05.030

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  3 in total

1.  Adherence of Staphylococcus aureus to Dyneema Purity® Patches and to Clinically Used Cardiovascular Prostheses.

Authors:  Amir Basir; Paul Gründeman; Frans Moll; Joost van Herwaarden; Gerard Pasterkamp; Reindert Nijland
Journal:  PLoS One       Date:  2016-09-01       Impact factor: 3.240

2.  Mechanical properties of all-suture anchors for rotator cuff repair.

Authors:  N S Nagra; N Zargar; R D J Smith; A J Carr
Journal:  Bone Joint Res       Date:  2017-02       Impact factor: 5.853

3.  Influence of Loading Conditions on the Mechanical Performance of Multifilament Coreless UHMWPE Sutures Used in Orthopaedic Surgery.

Authors:  Maria Prado-Novoa; Laura Perez-Sanchez; Belen Estebanez; Salvador Moreno-Vegas; Ana Perez-Blanca
Journal:  Materials (Basel)       Date:  2022-03-31       Impact factor: 3.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.