Literature DB >> 25104423

Fracture of the cobalt-chromium modular femoral neck component in total hip arthroplasty.

M-L Mencière1, T Amouyel2, J Taviaux2, M Bayle2, C Laterza2, P Mertl2.   

Abstract

Total hip prostheses using cervico-metaphyseal modularity were designed to better replicate the geometry of the native extra-medullary femur. However, they are associated with numerous complications including corrosion, disassembly, pseudotumours and, most notably, fractures of the modular neck. All reported cases of modular neck fractures occurred with titanium components (Ti-6Al-4V). To prevent this weakness, manufacturers developed modular necks made of cobalt-chromium (Co-Cr). We report a fracture of a long, 8° varus, Co-Cr modular neck connected to a 36-mm short (-3.5mm) femoral head. The fracture occurred 22 months post-implantation in a woman who had a low level of physical activity and a body mass index of 28.7 kg/m(2). To our knowledge, this case is the first reported instance of Co-Cr modular neck fracture. It may challenge the wisdom of further developing this modularity design, as our patient had none of the known risk factors for modular neck fracture. In addition, cases of pseudotumour have been reported with Co-Cr modular necks subjected to fretting corrosion, which contributed to the fracture in our patient.
Copyright © 2014 Elsevier Masson SAS. All rights reserved.

Entities:  

Keywords:  Cobalt-chromium; Corrosion; Fracture; Hip prosthesis; Modular neck

Mesh:

Substances:

Year:  2014        PMID: 25104423     DOI: 10.1016/j.otsr.2014.03.027

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  6 in total

1.  What Factors Are Associated With Neck Fracture in One Commonly Used Bimodular THA Design? A Multicenter, Nationwide Study in Slovenia.

Authors:  Simon Kovač; Blaž Mavčič; Marko Kotnik; Vesna Levašič; Mateja Sirše; Samo Karel Fokter
Journal:  Clin Orthop Relat Res       Date:  2019-06       Impact factor: 4.176

2.  Cementless modular neck stems: are they a safe option in primary total hip arthroplasty?

Authors:  José M Pelayo-de-Tomás; José L Rodrigo-Pérez; Carlos D Novoa-Parra; Alejandro Lizaur-Utrilla; María Morales-Suárez-Varela; José Antonio Blas-Dobón
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-10-29

3.  A New Case of Fracture of a Modular Femoral Neck Device After a Total Hip Arthroplasty.

Authors:  Klemens Trieb; Nicola Stadler
Journal:  Open Orthop J       Date:  2015-05-15

4.  High Risk of Failure With Bimodular Femoral Components in THA.

Authors:  Aidin Eslam Pour; Robert Borden; Takayuki Murayama; Mary Groll-Brown; J David Blaha
Journal:  Clin Orthop Relat Res       Date:  2015-09-01       Impact factor: 4.176

5.  Retrieval Findings of Recalled Dual-Taper Hips.

Authors:  Anna Di Laura; Harry S Hothi; Johann Henckel; Young-Min Kwon; John A Skinner; Alister J Hart
Journal:  J Bone Joint Surg Am       Date:  2018-10-03       Impact factor: 5.284

6.  Femoral Stem Fracture in Hip Revision Arthroplasty: A Systematic Literature Review of the Real-World Evidence.

Authors:  Matthias Sukopp; Dominic Taylor; Raimund Forst; Frank Seehaus
Journal:  Z Orthop Unfall       Date:  2021-04-13       Impact factor: 0.923

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.