An evaluation of the RapidHIT(®) system for reliably genotyping reference samples.

Bobby L LaRue1, Andrea Moore2, Jonathan L King2, Pamela L Marshall2, Bruce Budowle3.   

Abstract

Short tandem repeat (STR) typing is used routinely for associating or excluding individuals with biological evidence left at a crime scene. Improvements have been made to reduce the turnaround time and labor involved with profile generation, but there is still some lag time between sample collection and interpretation of results. The RapidHIT(®) (IntegenX; Pleasanton, CA, USA) system is an automated instrument that is configured to perform DNA extraction, bead-based DNA normalization, amplification, electrophoresis of PCR amplicons, and data analysis of five reference swabs simultaneously. The RapidHIT system provided reliable STR profiles from reference buccal swabs in approximately 90min with nominal "hands-on" sample loading time with no evidence of contamination between samples. The overall success rate of typing buccal swabs was comparable to standard typing systems. In the event of a failed run due to instrument failure, the swab can be removed from the cartridge and reanalyzed in the RapidHIT system or with standard STR genotyping workflows.
Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Automated STR typing; R-DNA; Rapid DNA; Rapid genotyping; STR

Mesh:

Year:  2014        PMID: 25086874     DOI: 10.1016/j.fsigen.2014.06.012

Source DB:  PubMed          Journal:  Forensic Sci Int Genet        ISSN: 1872-4973            Impact factor:   4.882


  3 in total

1.  Inexpensive, rapid prototyping of microfluidic devices using overhead transparencies and a laser print, cut and laminate fabrication method.

Authors:  Brandon L Thompson; Yiwen Ouyang; Gabriela R M Duarte; Emanuel Carrilho; Shannon T Krauss; James P Landers
Journal:  Nat Protoc       Date:  2015-05-14       Impact factor: 13.491

Review 2.  The future of forensic DNA analysis.

Authors:  John M Butler
Journal:  Philos Trans R Soc Lond B Biol Sci       Date:  2015-08-05       Impact factor: 6.237

3.  Validation of the Applied Biosystems RapidHIT ID instrument and ACE GlobalFiler Express sample cartridge.

Authors:  Jennifer Churchill Cihlar; Kapema Bupe Kapema; Bruce Budowle
Journal:  Int J Legal Med       Date:  2021-10-13       Impact factor: 2.686

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.