Motion tracking in the liver: validation of a method based on 4D ultrasound using a nonrigid registration technique.

Sinara Vijayan1, Stefan Klein2, Erlend Fagertun Hofstad3, Frank Lindseth4, Brynjulf Ystgaard5, Thomas Langø3.   

Abstract

PURPOSE: Treatments like radiotherapy and focused ultrasound in the abdomen require accurate motion tracking, in order to optimize dosage delivery to the target and minimize damage to critical structures and healthy tissues around the target. 4D ultrasound is a promising modality for motion tracking during such treatments. In this study, the authors evaluate the accuracy of motion tracking in the liver based on deformable registration of 4D ultrasound images.
METHODS: The offline analysis was performed using a nonrigid registration algorithm that was specifically designed for motion estimation from dynamic imaging data. The method registers the entire 4D image data sequence in a groupwise optimization fashion, thus avoiding a bias toward a specifically chosen reference time point. Three healthy volunteers were scanned over several breathing cycles (12 s) from three different positions and angles on the abdomen; a total of nine 4D scans for the three volunteers. Well-defined anatomic landmarks were manually annotated in all 96 time frames for assessment of the automatic algorithm. The error of the automatic motion estimation method was compared with interobserver variability. The authors also performed experiments to investigate the influence of parameters defining the deformation field flexibility and evaluated how well the method performed with a lower temporal resolution in order to establish the minimum frame rate required for accurate motion estimation.
RESULTS: The registration method estimated liver motion with an error of 1 mm (75% percentile over all datasets), which was lower than the interobserver variability of 1.4 mm. The results were only slightly dependent on the degrees of freedom of the deformation model. The registration error increased to 2.8 mm with an eight times lower temporal resolution.
CONCLUSIONS: The authors conclude that the methodology was able to accurately track the motion of the liver in the 4D ultrasound data. The authors believe that the method has potential in interventions on moving abdominal organs such as MR or ultrasound guided focused ultrasound therapy and radiotherapy, pending the method is enabled to run in real-time. The data and the annotations used for this study are made publicly available for those who would like to test other methods on 4D liver ultrasound data.

Entities:  

Mesh:

Year:  2014        PMID: 25086560     DOI: 10.1118/1.4890091

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  A 3D multi-modal intelligent intervention system using electromagnetic navigation for real-time positioning and ultrasound images: a prospective randomized controlled trial.

Authors:  Weiwei Tang; Yun Zhou; Hui Zhao; Guangshun Sun; Dawei Rong; Zhitao Li; Meng Hu; Liu Han; Xu He; Suming Zhao; Xiaoyang Chen; Zhongming Li; Hongxin Yuan; Songwang Chen; Qian Wang; Zhouxiao Li; Jianping Gu; Xuehao Wang; Jinhua Song
Journal:  Ann Transl Med       Date:  2022-06

2.  The 2014 liver ultrasound tracking benchmark.

Authors:  V De Luca; T Benz; S Kondo; L König; D Lübke; S Rothlübbers; O Somphone; S Allaire; M A Lediju Bell; D Y F Chung; A Cifor; C Grozea; M Günther; J Jenne; T Kipshagen; M Kowarschik; N Navab; J Rühaak; J Schwaab; C Tanner
Journal:  Phys Med Biol       Date:  2015-07-02       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.