Literature DB >> 25086522

Quality control procedures for dynamic treatment delivery techniques involving couch motion.

Victoria Y Yu1, Benjamin P Fahimian1, Lei Xing1, Dimitre H Hristov1.   

Abstract

In this study, the authors introduce and demonstrate quality control procedures for evaluating the geometric and dosimetric fidelity of dynamic treatment delivery techniques involving treatment couch motion synchronous with gantry and multileaf collimator (MLC). Tests were designed to evaluate positional accuracy, velocity constancy and accuracy for dynamic couch motion under a realistic weight load. A test evaluating the geometric accuracy of the system in delivering treatments over complex dynamic trajectories was also devised. Custom XML scripts that control the Varian TrueBeam™ STx (Serial #3) axes in Developer Mode were written to implement the delivery sequences for the tests. Delivered dose patterns were captured with radiographic film or the electronic portal imaging device. The couch translational accuracy in dynamic treatment mode was 0.01 cm. Rotational accuracy was within 0.3°, with 0.04 cm displacement of the rotational axis. Dose intensity profiles capturing the velocity constancy and accuracy for translations and rotation exhibited standard deviation and maximum deviations below 3%. For complex delivery involving MLC and couch motions, the overall translational accuracy for reproducing programmed patterns was within 0.06 cm. The authors conclude that in Developer Mode, TrueBeam™ is capable of delivering dynamic treatment delivery techniques involving couch motion with good geometric and dosimetric fidelity.

Entities:  

Mesh:

Year:  2014        PMID: 25086522     DOI: 10.1118/1.4886757

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Dosimetric accuracy of dynamic couch rotation during volumetric modulated arc therapy (DCR-VMAT) for primary brain tumours.

Authors:  Gregory Smyth; Philip M Evans; Jeffrey C Bamber; Henry C Mandeville; A Rollo Moore; Liam C Welsh; Frank H Saran; James L Bedford
Journal:  Phys Med Biol       Date:  2019-04-05       Impact factor: 3.609

2.  Comparison of conventional and advanced radiotherapy techniques for left-sided breast cancer after breast conserving surgery.

Authors:  Yibo Xie; Daniel Bourgeois; Beibei Guo; Rui Zhang
Journal:  Med Dosim       Date:  2020-07-07       Impact factor: 1.482

3.  The development and verification of a highly accurate collision prediction model for automated noncoplanar plan delivery.

Authors:  Victoria Y Yu; Angelia Tran; Dan Nguyen; Minsong Cao; Dan Ruan; Daniel A Low; Ke Sheng
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

Review 4.  Recent developments in non-coplanar radiotherapy.

Authors:  Gregory Smyth; Philip M Evans; Jeffrey C Bamber; James L Bedford
Journal:  Br J Radiol       Date:  2019-02-01       Impact factor: 3.039

5.  Equivalency of beam scan data collection using a 1D tank and automated couch movements to traditional 3D tank measurements.

Authors:  Nels C Knutson; Matthew C Schmidt; Matthew D Belley; Ngoc Nguyen; Michael Price; Sasa Mutic; Erno Sajo; H Harold Li
Journal:  J Appl Clin Med Phys       Date:  2018-09-06       Impact factor: 2.102

  5 in total