Constituent components of out-of-field scatter dose for 18-MV intensity modulated radiation therapy versus 3-dimensional conformal radiation therapy: a comparison with 6-MV and implications for carcinogenesis.

Jeremy D Ruben1, Ryan Smith2, Craig M Lancaster3, Matthew Haynes4, Phillip Jones4, Vanessa Panettieri4.   

Abstract

PURPOSE: To characterize and compare the components of out-of-field dose for 18-MV intensity modulated radiation therapy (IMRT) versus 3-dimensional conformal radiation therapy (3D-CRT) and their 6-MV counterparts and consider implications for second cancer induction. METHODS AND MATERIALS: Comparable plans for each technique/energy were delivered to a water phantom with a sloping wall; under full scatter conditions; with field edge abutting but outside the bath to prevent internal/phantom scatter; and with shielding below the linear accelerator head to attenuate head leakage. Neutron measurements were obtained from published studies.
RESULTS: Eighteen-megavolt IMRT produces 1.7 times more out-of-field scatter than 18-MV 3D-CRT. In absolute terms, however, differences are just approximately 0.1% of central axis dose. Eighteen-megavolt IMRT reduces internal/patient scatter by 13%, but collimator scatter (C) is 2.6 times greater than 18-MV 3D-CRT. Head leakage (L) is minimal. Increased out-of-field photon scatter from 18-MV IMRT carries out-of-field second cancer risks of approximately 0.2% over and above the 0.4% from 18-MV 3D-CRT. Greater photoneutron dose from 18-MV IMRT may result in further maximal, absolute increased risk to peripheral tissue of approximately 1.2% over 18-MV 3D-CRT. Out-of-field photon scatter remains comparable for the same modality irrespective of beam energy. Machine scatter (C+L) from 18 versus 6 MV is 1.2 times higher for IMRT and 1.8 times for 3D-CRT. It is 4 times higher for 6-MV IMRT versus 3D-CRT. Reduction in internal scatter with 18 MV versus 6 MV is 27% for 3D-CRT and 29% for IMRT. Compared with 6-MV 3D-CRT, 18-MV IMRT increases out-of-field second cancer risk by 0.2% from photons and adds 0.28-2.2% from neutrons.
CONCLUSIONS: Out-of-field photon dose seems to be independent of beam energy for both techniques. Eighteen-megavolt IMRT increases out-of-field scatter 1.7-fold over 3D-CRT because of greater collimator scatter despite reducing internal/patient scatter. Out-of-field carcinogenic risk is thus increased (but improved in-field dose conformity may offset this). Potentially increased carcinogenic risk should be weighed against any benefit 18-MV IMRT may provide.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 25084609     DOI: 10.1016/j.ijrobp.2014.05.052

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  3 in total

1.  Cost-effectiveness analysis of advanced radiotherapy techniques for post-mastectomy breast cancer patients.

Authors:  Yibo Xie; Beibei Guo; Rui Zhang
Journal:  Cost Eff Resour Alloc       Date:  2020-08-03

2.  Evaluation of surface and shallow depth dose reductions using a Superflab bolus during conventional and advanced external beam radiotherapy.

Authors:  Jihyung Yoon; Yibo Xie; Rui Zhang
Journal:  J Appl Clin Med Phys       Date:  2018-02-10       Impact factor: 2.102

3.  Clinical Characteristics and Predictive Outcomes of Recurrent Nasopharyngeal Carcinoma-A Lingering Pitfall of the Long Latency.

Authors:  Yung-Hsuan Chen; Sheng-Dean Luo; Shao-Chun Wu; Ching-Nung Wu; Tai-Jan Chiu; Yu-Ming Wang; Yao-Hsu Yang; Wei-Chih Chen
Journal:  Cancers (Basel)       Date:  2022-08-04       Impact factor: 6.575

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.