Accuracy and repeatability of 3 apex locators in locating root canal perforations: an ex vivo study.

Fábio Luiz Cunha D'Assunção1, Julio Cézar Nascimento Sousa2, Kayo César Amaro Felinto3, Thiago Clístines de Medeiros4, Diego Tavares Leite3, Raissa Bezerra de Lucena2, Joab de Oliveira Lima5.   

Abstract

INTRODUCTION: This study was an ex vivo evaluation of the accuracy and repeatability of 3 electronic apex locators (EALs) in locating simulated root canal perforations.
METHODS: Thirty-five human teeth were used. Perforations were placed 5 mm from the apex of each specimen. A #15 K-file was attached to a micrometer, and the specimens were fixed in a mounting model. The EALs were connected to the files, and electronic measurements of the simulated perforations were obtained. This procedure was repeated, the files were cemented in place, and x-rays were obtained to evaluate the distances between the measurement files and the simulated perforations. The obtained results were compared with the corresponding control lengths. Three investigators evaluated the distance of the file position in relation to the perforation for each specimen. The 3 investigators reached consensus regarding these measurements.
RESULTS: Statistical analyses of the results revealed that the EAL reliabilities in the exact detections of the root canal perforations were 97.1% for the Mini Apex Locator, 100% for the Root SW, and 91.4% for the Root ZX II. Analysis of variance revealed that the accuracies of the 3 devices were not different (P > .05). The results also indicated high levels of repeatability for each of the 3 EALs (intraclass correlation coefficients: 0998, 0.999, and 0.999 for the Mini Apex Locator, Root SW, and Root ZX II, respectively).
CONCLUSIONS: It was concluded that each of these 3 EALs provided excellent ex vivo accuracy in locating root perforations and that all tested devices possess excellent ex vivo repeatability.
Copyright © 2014 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Apex locators; Mini Apex Locator; Root ZX II; root perforation

Mesh:

Substances:

Year:  2014        PMID: 25069941     DOI: 10.1016/j.joen.2014.02.004

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  3 in total

1.  Comparative Evaluation of Accuracy of Different Apex Locators: Propex IQ, Raypex 6, Root ZX, and Apex ID with CBCT and Periapical Radiograph-In Vitro Study.

Authors:  Okba Mahmoud; Mawada Hassan Awad Abdelmagied; Ahmad Hisham Dandashi; Bakr Nssaief Jasim; Hussam Alddin Tawfik Kayali; Saaid Al Shehadat
Journal:  Int J Dent       Date:  2021-05-04

2.  Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants.

Authors:  Sakshi Bilaiya; Pallav Mahesh Patni; Pradeep Jain; Sanket Hans Pandey; Swadhin Raghuwanshi; Bhupesh Bagulkar
Journal:  Eur Endod J       Date:  2020-03-19

3.  Accuracy of two electronic apex locators in locating root perforations in curved canals in dry and wet conditions: A comparative in vitro study.

Authors:  Monisha Parshotam Khatri; Sheetal B Ghivari; Madhu Pujar; Reshma Faras; Pallavi Gopeshetti; Amulya Vanti
Journal:  Dent Res J (Isfahan)       Date:  2019-11-12
  3 in total