Marta Fiedoruk1, Elżbieta Mieczkowska1, Robert Koncki1, Lukasz Tymecki2.
Abstract
A miniature flow-through detector useful for bimodal, photometric and fluorimetric, determination of phosphates has been developed. This optoelectronic device made of four light emitting diodes (LEDs) integrated in the form of 85 µL optical cell is easily applied in flow analysis manifolds. These LEDs play the roles of light source for photometric measurements, fluorescence inductors and detector of absorbance and fluorescence. For photometric mode of determinations a phosphomolybdenum blue method has been applied. The fluorimetric method of phosphate determination is based on quenching of rhodamine fluorescence by the heteropolyacid. The developed detector used in a simple three-channel flow injection analysis (FIA) system allows photometric or fluorimetric determination of phosphate in the wide range of concentration. The detection limits found for photometric and fluorimetric modes of FIA measurements are 5.5 mg L(-1) and 10.4 µg L(-1), respectively. The potential utility of the flow-through detector for the needs of food and clinical analysis has been demonstrated.Entities:
Keywords: Flow analysis; Fluorimetry; Optoelectronic detector; Phosphate; Photometry
Mesh:
Substances:
Year: 2014 PMID: 25059150 DOI: 10.1016/j.talanta.2014.04.086
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057