| Literature DB >> 25054152 |
Xiaoni Li1, Hongshun Yang2, Hailiang Huang3, Tao Zhu1.
Abstract
Transwell Boyden chamber based migration/invasion assay is a simple and extensively used approach for the characterization of cell motility in vitro. Cell motility is quantified by counting the number of cells that pass through the filter membrane. The counting is usually performed manually, which is laborious and error prone. We have therefore developed CELLCOUNTER, an application that is capable of recognizing and counting the total number of cells through an intuitive graphical user interface. The counting can be performed in batch, and the counting results can be visualized and further curated manually. CELLCOUNTER will be helpful in streamlining the experimental process and improving the reliability of the data acquisition.Entities:
Mesh:
Year: 2014 PMID: 25054152 PMCID: PMC4099152 DOI: 10.1155/2014/863564
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1The graphical user interface of CELLCOUNTER. Green circles indicate automatically recognized cells and red circles indicate manually labeled cells. Functions for buttons/inputs are (1) manual counting/curation; (2) zooming in; (3) zooming out; (4) restoring to the original image size; (5) saving the counting results (including images); (6) setting the radius of the label (for visualization only); (7) setting the radius of small wells (in px, see the main text for more details, default recommended).
Images used in this study. Cells were stained with 0.1% crystal violet for 20 minutes for the deep staining group and 8 minutes for the light staining group. High density images refer to those having over 400 cells and low density images refer to those having less than 200 cells. Biology stimulation of cells was discussed in Section 2.
| ID | Cell line | Stimulation | Stain | Density | Number of cells | ||
|---|---|---|---|---|---|---|---|
| Expert | CELLCOUNTER | Beginner | |||||
| 1 | T47D | No | — | High | 931 | 957 | 955 |
| 2 | MCF-7 | No | Deep | High | 792 | 748 | 538 |
| 3 | MDA-MB-231 | No | — | High | 475 | 479 | 413 |
| 4 | pc-3 | No | — | — | 523 | 507 | 390 |
| 5 | MCF-7 | — | Deep | — | 314 | 322 | 263 |
| 6 | MCF-7 | Yes | Deep | — | 359 | 364 | 300 |
| 7 | MDA-MB-231 | Yes | — | High | 566 | 572 | 366 |
| 8 | Hep3B | — | Light | — | 284 | 279 | 249 |
| 9 | MDA-MB-435s | — | — | — | 464 | 461 | 430 |
| 10 | Hep3B | — | — | — | 186 | 178 | 144 |
| 11 | MCF-7 | — | Light | Low | 165 | 159 | — |
| 12 | MCF-7 | Yes | — | — | 403 | 403 | — |
| 13 | MCF-7 | No | Deep | — | 417 | 395 | — |
| 14 | MCF-7 | Yes | Light | — | 189 | 187 | — |
| 15 | MCF-7 | — | Light | Low | 94 | 98 | — |
| 16 | MCF-7 | Yes | — | Low | 120 | 118 | — |
| 17 | MCF-7 | — | — | Low | 139 | 140 | — |
| 18 | MCF-7 | — | Light | — | 8 | 8 | — |
| 19 | MCF-7 | — | Deep | High | 554 | 553 | — |
| 20 | MCF-7 | — | — | Low | 191 | 192 | — |
Figure 2Cell counting results for 10 assay images from 3 experts, CELLCOUNTER, and 3 beginners. Error bars are standard deviations (only available for human counting results).
Figure 3Counting results from CELLCOUNTER (a), CELLPROFILER (b), and IMAGEJ (c) comparing with human counting results.