Bingquan Zhu1, Yongbing Wang2, Guozheng Yan1, Pingping Jiang1, Zhiqiang Liu1.
Abstract
Electrical stimulation has been suggested as a possible treatment for various functional gastrointestinal disorders (FGID). This paper presents a transcutaneous power supplied implantable electrical stimulation system. This technology solves the problem of supplying extended power to an implanted electrical stimulator. After implantation, the stimulation parameters can be reprogrammed by the external controller and then transmitted to the implanted stimulator. This would enable parametric studies to investigate the efficacy of various stimulation parameters in promoting gastrointestinal contractions. A pressure detector in the internal stimulator can provide real-time feedback about variations in the gastrointestinal tract. An optimal stimulation protocol leading to cecal contractions has been proposed: stimulation bursts of 3 ms pulse width, 10 V amplitude, 40 Hz frequency, and 20 s duration. The animal experiment demonstrated the functionality of the system and validated the effects of different stimulation parameters on cecal contractions.Entities:
Year: 2014 PMID: 25053939 PMCID: PMC4099221 DOI: 10.1155/2014/728572
Source DB: PubMed Journal: Gastroenterol Res Pract ISSN: 1687-6121 Impact factor: 2.260
Figure 1System principle.
Figure 2Stimulation pulses.
Stimulation parameters.
| Parameters | Parameters' range | |
|---|---|---|
| Unit | Range | |
| Amplitude (peak-peak) | V | 8–25 |
| Pulse cycle | ms | 0.1–999 |
| Pulse width | ms | 0.1–999 |
| Stimulation duration | s | 1–999 |
Figure 3Power transmission link model.
Figure 4Schematic of the transcutaneous power supply system.
Parameters of transmitting coil and receiving coil.
| Parameters | Radius (mm) | Height (mm) | Inductance ( | Capacitance (nF) | Impedance (Ω) |
|
|---|---|---|---|---|---|---|
| Transmitting coil | 60 | 5.6 | 251.0 | 1.9 | 2.0 | 235.0 |
| Receiving coil | 35 | 3 | 140.0 | 3.4 | 4.6 | 44.0 |
Figure 5Test of transcutaneous power supply system. Test bench (a) and results in axial distance (b), radial distance (c), and radial angle (d).
Figure 6Animal experiment.
Reactions of the cecum to different stimulation parameters.
| Stimulation parameters | Healthy pig | STC pig | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
| Incubation | Myoelectrical | Pressure rise | Shrinkage | Incubation | Myoelectrical | Pressure rise | Shrinkage |
| 7.5 | 1 | 10 | 16.1 ± 2.1 | 6.2 ± 1.8 | 5.2 ± 1.1 | 8.2 ± 2.26 | 21.1 ± 2.3 | 2.1 ± 0.4 | 3.2 ± 0.5 | 7.3 ± 1.19 |
| 7.5 | 3 | 10 | 15.2 ± 3.5 | 14.1 ± 3.2 | 13.6 ± 2.7 | 14.5 ± 3.15 | 19.8 ± 3.7 | 5.4 ± 1.6 | 6.6 ± 2.1 | 6.4 ± 1.41 |
| 7.5 | 5 | 10 | 14.1 ± 2.4 | 22.9 ± 4.1 | 24.1 ± 2.3 | 18.7 ± 3.71 | 18.8 ± 3.5 | 8.3 ± 1.2 | 13.9 ± 0.8 | 12.6 ± 2.48 |
| 10 | 0.1 | 40 | 9.5 ± 1.2 | 8.3 ± 1.5 | 9.8 ± 1.1 | 11.4 ± 2.49 | 15.1 ± 2.5 | 4.8 ± 1.2 | 5.2 ± 1.0 | 4.4 ± 1.52 |
| 10 | 0.3 | 40 | 9.3 ± 2.1 | 12.1 ± 2.8 | 9.4 ± 1.4 | 15.7 ± 2.41 | 14.6 ± 1.9 | 7.5 ± 2.3 | 3.7 ± 0.6 | 8.3 ± 2.09 |
| 10 | 1 | 40 | 9.0 ± 1.6 | 27.8 ± 4.3 | 12.6 ± 2.4 | 15.3 ± 4.29 | 12.1 ± 1.1 | 13.4 ± 1.7 | 5.5 ± 0.7 | 11.6 ± 2.84 |
| 10 | 3 | 40 | 8.8 ± 0.9 | 29.2 ± 7.8 | 27.5 ± 3.4 | 31.4 ± 4.44 | 13.4 ± 0.9 | 15.3 ± 1.7 | 22.8 ± 5.0 | 29.4 ± 3.82 |
| 10 | 5 | 40 | 8.3 ± 1.1 | 36.5 ± 11.2 | 37.4 ± 4.2 | 38.4 ± 6.56 | 12.5 ± 1.1 | 18.5 ± 3.1 | 36.0 ± 5.4 | 32.4 ± 4.72 |
| 10 | 10 | 40 | 7.9 ± 0.8 | 47.6 ± 7.5 | 59.5 ± 4.7 | 40.7 ± 7.01 | 11.0 ± 1.4 | 37.5 ± 8.1 | 44.1 ± 2.9 | 39.4 ± 8.82 |
| 12 | 0.1 | 120 | 5.1 ± 0.8 | 11.7 ± 1.8 | 20.0 ± 1.7 | 14.6 ± 4.32 | 12.5 ± 0.8 | 12 ± 0.5 | 12.6 ± 1.9 | 9.3 ± 2.19 |
| 12 | 0.3 | 120 | 4.2 ± 0.6 | 17.3 ± 2.1 | 34.2 ± 2.3 | 21.9 ± 6.17 | 9.3 ± 0.8 | 11.4 ± 1.3 | 12.8 ± 1.1 | 11.4 ± 2.42 |
| 12 | 1 | 120 | 4.1 ± 0.4 | 38.3 ± 4.2 | 51.9 ± 4.7 | 42.7 ± 8.81 | 8.8 ± 1.3 | 18.5 ± 2.1 | 36.1 ± 2.1 | 36.7 ± 7.01 |
| 12 | 3 | 120 | 3.8 ± 0.3 | 41.7 ± 5.1 | 60.1 ± 5.5 | 44.4 ± 9.32 | 8.9 ± 2.1 | 31.3 ± 7.9 | 43.2 ± 3.0 | 38.5 ± 11.75 |
| 12 | 5 | 120 | 3.6 ± 0.2 | Tetanic contraction | 71.4 ± 1.3 | 46.8 ± 11.04 | 5.6 ± 0.8 | 43.7 ± 11.2 | 43.1 ± 5.3 | 39.6 ± 12.02 |
*The unit of myoelectrical activity is times per 5 min which is counted after the stimulation.
Figure 7Cecal pressure of an STC pig under different stimulation parameters.
Figure 8Implantation of the receiving coil.