Hun Lee1, Kwanghyun Lee1, Ji Min Ahn1, Eung Kweon Kim2, Bradford Sgrignoli1, Tae-im Kim3.
Abstract
PURPOSE: We investigated the correlations between optical quality parameters obtained from the double-pass system and ocular aberrations obtained from the ray-tracing aberrometer in multifocal intraocular lens (IOL) implanted eyes.Entities:
Keywords: Double-pass system; cataract surgery; multifocal intraocular lens; optical quality; ray-tracing aberrometer
Mesh:
Year: 2014 PMID: 25048505 PMCID: PMC4108832 DOI: 10.3349/ymj.2014.55.5.1413
Source DB: PubMed Journal: Yonsei Med J ISSN: 0513-5796 Impact factor: 2.759
Postoperative Data
CDVA, corrected distance visual acuity; logMAR, logarithm of the minimum angle of resolution; MTF, modulation transfer function; RMS, root mean square.
Correlations between Parameters Obtained from the Double-Pass System (MTF Cutoff Frequency, Strehl Ratio, and Objective Pseudo-Accommodation) and Ocular Aberrations Obtained from the Ray-Tracing Aberrometer
MTF, modulation transfer function; RMS, root mean square.
*Significant p-value (p<0.05 is considered statistically significant).
†p-value is from Spearman's correlation analysis.
‡Objective scatter index was measured from the double-pass system.
Correlations between Parameters Obtained from the Double-Pass System (MTF Cutoff Frequency, Strehl Ratio, and Objective Pseudo-Accommodation) and Manifest Refraction Values or Subjective Visual Acuity
MTF, modulation transfer function; logMAR, logarithm of the minimum angle of resolution; CDVA, corrected distance visual acuity; UCDVA, uncorrected distance visual acuity.
*p-value is from Spearman's correlation analysis.
Correlations between Parameters Obtained from the Ray-Tracing Aberrometer (RMS Total Aberration, RMS Higher-Order Aberration, and Spherical Aberration) and Manifest Refraction Values or Subjective Visual Acuity
RMS, root mean square; logMAR, logarithm of the minimum angle of resolution; CDVA, corrected distance visual acuity; UCDVA, uncorrected distance visual acuity.
*p-value is from Spearman's correlation analysis.