Literature DB >> 25040659

A novel flow-based assay reveals discrepancies in ADAMTS-13 inhibitor assessment as compared with a conventional clinical static assay.

R Grillberger1, B Gruber, S Skalicky, G Schrenk, P Knöbl, B Plaimauer, P L Turecek, F Scheiflinger, H Rottensteiner.   

Abstract

BACKGROUND: Several static Bethesda-type assays are routinely used to determine ADAMTS-13-neutralizing autoantibodies in acquired thrombotic thrombocytopenic purpura (TTP), but the inhibitory activity of these antibodies has not been thoroughly evaluated under the more physiologic condition of flow.
OBJECTIVES: We investigated whether ADAMTS-13 inhibitor assessment with the FRETS-VWF73 assay is predictive for evaluation under flow.
METHODS: Anti-ADAMTS-13 autoantibodies were purified from patients with acquired TTP by chromatography involving an ADAMTS-13 affinity matrix and/or protein G. ADAMTS-13 activity was measured with the FRETS-VWF73 assay and a novel flow assay determining the ADAMTS-13-mediated decrease in platelet aggregate surface coverage, caused by perfusion of a suspension containing platelets, erythrocytes and von Willebrand factor (VWF) over a surface coated with extracellular matrix components. The neutralizing activities of ADAMTS-13 inhibitors were compared under static conditions and under flow by use of the two assays.
RESULTS: The suitability of the flow-based ADAMTS-13 activity assay for quantification of ADAMTS-13 inhibitors could be demonstrated by reversibility of the ADAMTS-13-dependent decrease in surface coverage upon addition of goat ADAMTS-13 antiserum. Testing the neutralizing activity of purified autoantibodies from six patients in the flow assay according to their FRETS-VWF73-based inhibitor titers gave rise to vastly different inhibitory effects, indicating a discrepancy in inhibitor assessment between static and flow conditions.
CONCLUSIONS: Anti-ADAMTS-13 autoantibodies may show inhibitory properties in vivo that are not consistent with the ADAMTS-13 inhibitor levels determined in routine static assays, possibly because certain epitopes are selectively exposed under shear. Consequently, the course of disease and treatment efficacy may vary among TTP patients, despite common inhibitor titers.
© 2014 International Society on Thrombosis and Haemostasis.

Entities:  

Keywords:  ADAMTS13 protein, human; autoantibodies; mechanical stress; thrombotic thrombocytopenic purpura; von Willebrand factor

Mesh:

Substances:

Year:  2014        PMID: 25040659     DOI: 10.1111/jth.12653

Source DB:  PubMed          Journal:  J Thromb Haemost        ISSN: 1538-7836            Impact factor:   5.824


  2 in total

1.  ADAMTS13 and 15 are not regulated by the full length and N-terminal domain forms of TIMP-1, -2, -3 and -4.

Authors:  Cenqi Guo; Anastasia Tsigkou; Meng Huee Lee
Journal:  Biomed Rep       Date:  2015-10-30

2.  Preserved Expression of mRNA Coding von Willebrand Factor-Cleaving Protease ADAMTS13 by Selenite and Activated Protein C.

Authors:  Michael L Ekaney; Clemens L Bockmeyer; Maik Sossdorf; Philipp A Reuken; Florian Conradi; Tobias Schuerholz; Markus F Blaess; Scott L Friedman; Wolfgang Lösche; Michael Bauer; Ralf A Claus
Journal:  Mol Med       Date:  2015-04-03       Impact factor: 6.354

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.