Validation of a short-scan-time imaging protocol for thallium-201 myocardial SPECT with a multifocal collimator.

Yoriko Horiguchi1, Tomohiro Ueda, Tomofumi Shiomori, Masahiko Kanna, Hirooki Matsushita, Tomoko Kawaminami, Yuta Sudo, Shinnosuke Kikuchi, Ryo Sasaki, Jun Hoshimiya, Yukiko Morita.   

Abstract

OBJECTIVE: IQ-SPECT (Siemens AG, Munich, Germany) is a highly sensitive single-photon-emission computed tomography (SPECT) myocardial perfusion imaging (MPI) system that uses a multifocal collimator. We searched for a suitable protocol for short-time imaging by IQ-SPECT in thallium-201 (Tl-201) MPI by evaluating phantom images and also by comparing human IQ-SPECT images with conventional SPECT images as reference standards.
METHODS: We assessed the image quality using the normalized mean square error (NMSE) and drew up count profiles in Tl-201 SPECT images acquired with IQ-SPECT in a phantom study. We also performed Tl-201 stress myocardial SPECT/CT in 21 patients and compared delayed images acquired by using IQ-SPECT with 36 or 17 views per head with images obtained by using conventional SPECT.
RESULTS: The NMSE of SPECT images from IQ-SPECT with 36 views was approximately one-fifth of that with 17 views. The myocardial count profile of images with 17 views was lower than those of images with 36 or 104 views in some regions. Defect scores were significantly lower, and image quality scores higher, in images from conventional SPECT than in those from IQ-SPECT with 17 views. Defect scores and image quality scores were equivalent in images from conventional SPECT and those from IQ-SPECT with 36 views. Agreement with the results of conventional SPECT in terms of coronary artery territory-based defect judgment was the best in IQ-SPECT with 36 views with computed tomography-derived attenuation correction (CTAC): the kappa values for IQ-SPECT with 36 views were 0.76 (without CTAC) and 0.83 (with CTAC), and those for IQ-SPECT with 17 views were 0.62 (without CTAC) and 0.59 (with CTAC). The difference in quantitative tracer uptake between conventional SPECT images and IQ-SPECT images was significantly greater for IQ-SPECT images with 17 views than for those with 36 views.
CONCLUSIONS: Scanning with 36 views per head with CTAC may be appropriate for Tl-201 MPI using IQ-SPECT, because it provides images equivalent to those using conventional SPECT.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25038906     DOI: 10.1007/s12149-014-0880-3

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  7 in total

1.  How to use cardiac IQ•SPECT routinely? An overview of tips and tricks from practical experience to the literature.

Authors:  Eric Gremillet; Denis Agostini
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-04       Impact factor: 9.236

2.  Creation and characterization of normal myocardial perfusion imaging databases using the IQ·SPECT system.

Authors:  Koichi Okuda; Kenichi Nakajima; Shinro Matsuo; Chisato Kondo; Masayoshi Sarai; Yoriko Horiguchi; Takahiro Konishi; Masahisa Onoguchi; Takeshi Shimizu; Seigo Kinuya
Journal:  J Nucl Cardiol       Date:  2017-01-03       Impact factor: 5.952

3.  New techniques, distinctive population, unique normal databases.

Authors:  James R Galt
Journal:  J Nucl Cardiol       Date:  2017-05-04       Impact factor: 5.952

4.  Feasibility of one-eighth time gated myocardial perfusion SPECT functional imaging using IQ-SPECT.

Authors:  Federico Caobelli; James T Thackeray; Alberto Soffientini; Frank M Bengel; Claudio Pizzocaro; Ugo Paolo Guerra
Journal:  Eur J Nucl Med Mol Imaging       Date:  2015-08-01       Impact factor: 9.236

Review 5.  Normal values and standardization of parameters in nuclear cardiology: Japanese Society of Nuclear Medicine working group database.

Authors:  Kenichi Nakajima; Naoya Matsumoto; Tokuo Kasai; Shinro Matsuo; Keisuke Kiso; Koichi Okuda
Journal:  Ann Nucl Med       Date:  2016-02-20       Impact factor: 2.668

Review 6.  IQ·SPECT technology and its clinical applications using multicenter normal databases.

Authors:  Kenichi Nakajima; Koichi Okuda; Mitsuru Momose; Shinro Matsuo; Chisato Kondo; Masayoshi Sarai; Takayuki Shibutani; Masahisa Onoguchi; Takeshi Shimizu; A Hans Vija
Journal:  Ann Nucl Med       Date:  2017-09-22       Impact factor: 2.668

7.  Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR.

Authors:  Eero Hippeläinen; Teemu Mäkelä; Touko Kaasalainen; Erna Kaleva
Journal:  EJNMMI Phys       Date:  2017-07-19
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.