Naoki Nakano1, Tomoyuki Matsumoto, Hirotsugu Muratsu, Takehiko Matsushita, Koji Takayama, Tokio Matsuzaki, Ryosuke Kuroda, Masahiro Kurosaka.
Abstract
The present study aimed to examine the results of total knee arthroplasty (TKA) with the NexGen Legacy posterior-stabilized (LPS)-Flex system for osteoarthritis in the valgus knee. Between 2003 and 2011, 27 valgus knees in 26 patients who underwent TKA with the NexGen LPS-Flex implant were clinically and radiologically evaluated after a minimum follow-up of 2 years. The original diagnosis was osteoarthritis. Knee Society Knee Score (KSKS), Knee Society Function Score (KSFS), maximum flexion angle, maximum extension angle, and radiological femorotibial angle (FTA) were evaluated pre- and postoperatively. The mean KSKS improved from 42.6 before surgery to 82.2 by the final follow-up (p < 0.01), and the mean KSFS improved from 41.1 to 80.9 (p < 0.01). The mean maximum flexion angle changed from 109.1° to 117.3° (no statistical significance) and the maximum extension angle improved from -9.7° to -3.6° (p < 0.05) postoperatively. The postoperative radiological FTA was 172.4°, which was closer to a neutral angle (174°) than the preoperative FTA (166.4°) (p < 0.01). None of the patients had undergone revision surgery by the final follow-up. As a conclusion, TKA with the NexGen LPS-Flex implant for osteoarthritis in the valgus knee resulted in satisfactory improvement regarding objective outcomes, although a longer term follow-up with a greater number of cases is necessary to verify these results.Entities:
Mesh:
Year: 2014 PMID: 25031206 DOI: 10.1007/s00590-014-1505-1
Source DB: PubMed Journal: Eur J Orthop Surg Traumatol ISSN: 1633-8065