OBJECTIVES: Despite the recent success regarding the transplantation of tissue-engineered airways, the mechanical properties of these grafts are not well understood. Mechanical assessment of a tissue-engineered airway graft before implantation may be used in the future as a predictor of function. The aim of this preliminary work was to develop a noninvasive image-processing environment for the assessment of airway mechanics.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25029584 PMCID: PMC4081877 DOI: 10.6061/clinics/2014(07)11
Source DB: PubMed Journal: Clinics (Sao Paulo) ISSN: 1807-5932 Impact factor: 2.365
Figure 1Environment used for the measurement of airway respiratory mechanics using image processing. (a) Trachea immersed in Krebs-Henseleit solution in an open-top polyurethane conditioner. (b) Overview of the stereomicroscope, airway conditioner, pressure transducers, Fleisch pneumotachograph, and SAV.
Figure 2A trachea from the CONTROL group. (a) Normal image. (b) Segmented image. (c) Division of the areas of analysis.
Control group measurements.
| Measurement | A | K | KMCh |
| Pressure (cmH2O) | 71.8±1.8 | 72.5±1.7 | 73.4±1.5 |
| Flow (mL/s) | 13.1±0.2 | 13.0±0.1 | 13.0±0.2 |
| Resistance (cmH2O/mL·s-1) | 5.5±0.1 | 5.6±0.2 | 5.7±0.2 |
| Pulsation – distal (%) | 5.1%±2.1% | 5.4%±2.1% | 5.7%±1.8% |
| Pulsation – medial (%) | 1.5%±1.1% | 2.3%±0.8% | 2.4%±1.0% |
| Pulsation – proximal (%) | 0.6%±0.2% | 1.4%±0.6% | 1.9%±0.9% |
DECEL group measurements.
| Measurement | A | K | KMCh |
| Pressure (cmH2O) | 62.6±3.3 | 64.0±3.1 | 60.7±3.8 |
| Flow (mL/s) | 13.1±1.1 | 13.0±0.5 | 12.7±1.0 |
| Resistance (cmH2O/mL·s-1) | 4.9±0.3 | 5.0±0.2 | 4.9±0.3 |
| Pulsation – distal (%) | 4.2%±1.7% | 4.3%±1.8% | 4.4%±1.7% |
| Pulsation – medial (%) | 3.2%±1.6% | 2.9%±1.4% | 2.9%±1.4% |
| Pulsation – proximal (%) | 2.7%±2.0% | 2.7%±1.9% | 2.8%±2.2% |
RECEL group measurements.
| Measurement | A | K | KMCh |
| Pressure (cmH2O) | 61.2±1.8 | 61.0±2.0 | 61.9±2.2 |
| Flow (mL/s) | 14.3±1.2 | 12.9±0.3 | 13.3±0.5 |
| Resistance (cmH2O/mL·s-1) | 4.38±0.3 | 4.73±0.1 | 4.70±0.2 |
| Pulsation – distal (%) | 9.6%±7.9% | 12.5%±9.2% | 9.3%±7.0% |
| Pulsation – medial (%) | 8.0%±3.7% | 8.2%±3.4% | 8.3%±3.6% |
| Pulsation – proximal (%) | 5.6%±2.6% | 5.7%±2.6% | 5.8%±2.6% |