Literature DB >> 25025301

Diagnostic features of retinal nerve fiber layer rotation in skew deviation using optical coherence tomography.

John J Chen1, Randy H Kardon, Reid A Longmuir.   

Abstract

A 41-year-old woman with skew deviation had cyclotorsion of both eyes. This resulted in a falsely low probability plot of retinal nerve fiber layer thickness in adjacent clock hours on optical coherence tomography (OCT) due to displacement of the retinal nerve fiber layer peaks. Ocular cyclotorsion may cause misinterpretation of OCT probability plots. OCT retinal nerve fiber layer plots also may be used to objectively quantify the degree of ocular cyclotorsion.

Entities:  

Mesh:

Year:  2014        PMID: 25025301     DOI: 10.1097/WNO.0000000000000152

Source DB:  PubMed          Journal:  J Neuroophthalmol        ISSN: 1070-8022            Impact factor:   3.042


  2 in total

Review 1.  Avoiding Clinical Misinterpretation and Artifacts of Optical Coherence Tomography Analysis of the Optic Nerve, Retinal Nerve Fiber Layer, and Ganglion Cell Layer.

Authors:  John J Chen; Randy H Kardon
Journal:  J Neuroophthalmol       Date:  2016-12       Impact factor: 3.042

2.  The OCT RNFL Probability Map and Artifacts Resembling Glaucomatous Damage.

Authors:  Sol La Bruna; Anvit Rai; Grace Mao; Jennifer Kerr; Heer Amin; Zane Z Zemborain; Ari Leshno; Emmanouil Tsamis; Carlos Gustavo De Moraes; Donald C Hood
Journal:  Transl Vis Sci Technol       Date:  2022-03-02       Impact factor: 3.283

  2 in total