Investigation of a fatigue failure in a stainless steel femoral plate.

J B Marcomini1, C A R P Baptista2, J P Pascon2, R L Teixeira3, F P Reis2.   

Abstract

Surgical implants are exposed to severe working conditions and therefore a wide range of failure mechanisms may occur, including fatigue, corrosion, wear, fretting and combinations of them. The mechanical failures of metallic implants may also be influenced by several other factors, including the design, material, manufacturing, installation, postoperative complications and misuse. An 83-year-old patient suffered an oblique femoral shaft fracture due to a fall at home. A stainless steel locking compression plate (LCP) employed in the fracture reduction failed after four months and was sent back to the producer. A second LCP of the same type was implanted and also failed after six months. A failure analysis of the second femoral LCP is performed in this paper. The results demonstrate that poor material quality was decisive to the failure. The chemical analysis revealed a high P content in the steel, which is not in accordance to the standards. A combination of factors lead to LCP fracture and these include: brittle crack initiation due to phosphorus, segregation at grain boundaries, crack propagation due to cyclic loading and final fast fracture favored by the loss of ductility due to cold work.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Failure analysis; Fatigue; Femoral plate; Orthopaedic implants; Stainless steel

Mesh:

Substances:

Year:  2014        PMID: 25023519     DOI: 10.1016/j.jmbbm.2014.06.011

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  3 in total

1.  Finite element analysis of bone and implant stresses for customized 3D-printed orthopaedic implants in fracture fixation.

Authors:  Lina Yan; Joel Louis Lim; Jun Wei Lee; Clement Shi Hao Tia; Gavin Kane O'Neill; Desmond Y R Chong
Journal:  Med Biol Eng Comput       Date:  2020-02-19       Impact factor: 2.602

2.  A systematic review of the use of titanium versus stainless steel implants for fracture fixation.

Authors:  Collin C Barber; Matthew Burnham; Ogaga Ojameruaye; Michael D McKee
Journal:  OTA Int       Date:  2021-08-18

3.  Microstructure and Mechanical Properties of Modified 316L Stainless Steel Alloy for Biomedical Applications Using Powder Metallurgy.

Authors:  Sadaqat Ali; Muhammad Irfan; Usama Muhammad Niazi; Ahmad Majdi Abdul Rani; Ahmad Rashedi; Saifur Rahman; Muhammad Kamal Asif Khan; Mabkhoot A Alsaiari; Stanislaw Legutko; Jana Petrů; Antonin Trefil
Journal:  Materials (Basel)       Date:  2022-04-12       Impact factor: 3.748