| Literature DB >> 25015906 |
S-H Kang1, J-W Lee2, S-H Lim2, Y-H Kim2, M-K Kim3.
Abstract
The goal of the present study was to compare the accuracy of dental image replacement on a cone beam computed tomography (CBCT) image using digital image data from three-dimensional (3D) optical scanning of a dental cast, occlusal bite, and bite tray impression. A Bracket Typodont dental model was used. CBCT of the dental model was performed and the data were converted to stereolithography (STL) format. Three experimental materials, a dental cast, occlusal bite, and bite tray impression, were optically scanned in 3D. STL files converted from the CBCT of the Typodont model and the 3D optical-scanned STL files of the study materials were image-registered. The error range of each methodology was measured and compared with a 3D optical scan of the Typodont. For the three materials, the smallest error observed was 0.099±0.114mm (mean error±standard deviation) for registering the 3D optical scan image of the dental cast onto the CBCT dental image. Although producing a dental cast can be laborious, the study results indicate that it is the preferred method. In addition, an occlusal bite is recommended when bite impression materials are used.Keywords: computed tomography; dental occlusion; registration
Mesh:
Substances:
Year: 2014 PMID: 25015906 DOI: 10.1016/j.ijom.2014.06.009
Source DB: PubMed Journal: Int J Oral Maxillofac Surg ISSN: 0901-5027 Impact factor: 2.789