GlideScope videolaryngoscope versus flexible fiberoptic bronchoscope for awake intubation of morbidly obese patient with predicted difficult intubation.

Ashraf Abualhasan Abdellatif, Monaz Abdulrahman Ali.   

Abstract

BACKGROUND: Awake fiberoptic intubation is the gold standard for management of predicted difficult intubation. The purpose of this study was to test whether Glide Scope video laryngoscopy (GVL) will provide significant advantages over fiberoptic bronchoscopy (FOB) for awake intubation in morbidly obese patients with predicted difficult intubation. We therefore tested the hypothesis that intubation using GVL is faster than intubation with FOB.
METHODS: 64 morbidly obese patients with predicted difficult intubation undergoing laparoscopic bariatric surgery were enrolled in this study. Patients were randomly assigned to receive awake oral intubation by either GVL or FOB. After airway topical anesthesia and sedation using target controlled remifentanil infusion to a Ramsay sedation scale of 3, wee compared the two devices for time to intubate, successful intubation on first attempt, glottic view using Cormack and Lehane score system, response of the patient to scope, patients satisfaction and incidence of postoperative sore throat and hoarseness.
RESULTS: Intubation time was 84 +/- 37.9 seconds and 73.6 +/- 31.1 seconds for FOB and GVL respectively. 75% of patients were successfully intubated on the first attempt with FOB compared to 80.6% with GVL. Grade I/II glottic view was reported with GVL in 96.7% of patients compared to 100% with FOB. The highest target concentration of remifentanil to maintain patients sedated during intubation was 2.4 +/- 0.6 ng/ml and 2.2 +/- 0.8 ng/ml in FOB and GVL respectively. No significant differences regarding maximum patient response to intubation, adverse effects or patient satisfaction were recorded between groups.
CONCLUSION: GVL can be used as a useful alternative to FOB in morbidly obese patients with predicted difficult intubation.

Entities:  

Mesh:

Year:  2014        PMID: 25007692

Source DB:  PubMed          Journal:  Middle East J Anaesthesiol        ISSN: 0544-0440


  12 in total

1.  Supraglottic jet oxygenation and ventilation-assisted fibre-optic bronchoscope intubation in patients with difficult airways.

Authors:  Caineng Wu; Jianqi Wei; Qingyun Cen; Xuefan Sha; Qingxiang Cai; Wuhua Ma; Ying Cao
Journal:  Intern Emerg Med       Date:  2016-09-14       Impact factor: 3.397

2.  Nasotracheal intubation of a patient with restricted mouth opening using a McGrath MAC X-Blade and Magill forceps.

Authors:  Zehra İpek Arslan; P Ozdal; D Ozdamar; H Agır; M Solak
Journal:  J Anesth       Date:  2016-06-24       Impact factor: 2.078

3.  Laryngoscope and a new tracheal tube assist lightwand intubation in difficult airways due to unstable cervical spine.

Authors:  Cai-neng Wu; Wu-hua Ma; Jian-qi Wei; Hua-feng Wei; Qing-yun Cen; Qing-xiang Cai; Ying Cao
Journal:  PLoS One       Date:  2015-03-24       Impact factor: 3.240

4.  Anesthetic Management for Laser Excision of Ball-Valving Laryngeal Masses.

Authors:  Benjamin B Bruins; Natasha Mirza; Ernest Gomez; Joshua H Atkins
Journal:  Case Rep Anesthesiol       Date:  2015-05-24

5.  Comparative randomised study of GlideScope® video laryngoscope versus flexible fibre-optic bronchoscope for awake nasal intubation of oropharyngeal cancer patients with anticipated difficult intubation.

Authors:  Essam Abd El-Halim Mahran; Mohamed Elsayed Hassan
Journal:  Indian J Anaesth       Date:  2016-12

Review 6.  Recent trends in airway management.

Authors:  Joelle Karlik; Michael Aziz
Journal:  F1000Res       Date:  2017-02-17

7.  Videolaryngoscopy versus fiberoptic bronchoscope for awake intubation - a systematic review and meta-analysis of randomized controlled trials.

Authors:  Jia Jiang; Da-Xu Ma; Bo Li; An-Shi Wu; Fu-Shan Xue
Journal:  Ther Clin Risk Manag       Date:  2018-10-15       Impact factor: 2.423

8.  Supraglottic airway device versus a channeled or non-channeled blade-type videolaryngoscope for accidental extubation in the prone position: A randomized crossover manikin study.

Authors:  Hiroyuki Oshika; Yukihide Koyama; Masataka Taguri; Koichi Maruyama; Go Hirabayashi; Shoko Merrit Yamada; Masashi Kohno; Tomio Andoh
Journal:  Medicine (Baltimore)       Date:  2018-06       Impact factor: 1.889

9.  Assessment of awake i-gel™ insertion for fiberoptic-guided intubation in patients with predicted difficult airway: A prospective, observational study.

Authors:  Julian Arevalo Ludeña; Jose Juan Arcas Bellas; Rafael Alvarez Rementeria; Luis Enrique Muñoz Alameda
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2018 Oct-Dec

10.  Apneic laryngeal oxygenation during elective fiberoptic intubation - a technical simulation.

Authors:  Daniel C Schroeder; Wolfgang A Wetsch; Simon-Richard Finke; Fabian Dusse; Bernd W Böttiger; Holger Herff
Journal:  BMC Anesthesiol       Date:  2020-12-09       Impact factor: 2.217

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.