S F Karg1, V Troncale, U Drechsler, P Mensch, P Das Kanungo, H Schmid, V Schmidt, L Gignac, H Riel, B Gotsmann.
Abstract
Precise measurements of a complete set of thermoelectric parameters on a single indium-arsenide nanowire (NW) have been performed using highly sensitive, micro-fabricated sensing devices based on the heater/sensor principle. The devices were fabricated as micro electro-mechanical systems consisting of silicon nitride membranes structured with resistive gold heaters/sensors. Preparation, operation and characterization of the devices are described in detail. Thermal decoupling of the heater/sensor platforms has been optimized reaching thermal conductances as low as 20 nW K(-1) with a measurements sensitivity below 20 nW K(-1). The InAs NWs were characterized in terms of thermal conductance, four-probe electrical conductance and thermopower (Seebeck coefficient), all measured on a single NW. The temperature dependence of the parameters determining the thermoelectric figure-of-merit of an InAs NW was acquired in the range 200-350 K featuring a minor decrease of the thermal conductivity from 2.7 W (m K)(-1) to 2.3 W (m K)(-1).Entities:
Year: 2014 PMID: 25004861 DOI: 10.1088/0957-4484/25/30/305702
Source DB: PubMed Journal: Nanotechnology ISSN: 0957-4484 Impact factor: 3.874