Literature DB >> 25002315

[Pneumatic tube system for transport of laboratory samples: preanalytical aspects].

Judit Tóth1, Ágota Lenkey1, Anna V Oláh1, Julianna Köteles1, Valéria Kissné Sziráki1, Adrienne Kerényi1, János Kappelmayer1.   

Abstract

INTRODUCTION: A considerable proportion of laboratory errors occurs in the preanalytical phase. AIM: The aims of the authors were to study preanalytical errors in routine and emergency laboratory diagnostics in a regional clinical laboratory and evaluate the effect of the pneumatic tube system on turnaround time and laboratory results.
METHOD: The ratio of preanalytical errors and reasons of test rejection were analysed. In addition, the effects of pneumatic tube and manual transport on the occurrence of hemolysis and platelet activation were compared.
RESULTS: Using the pneumatic tube transport system, preanalytical error was below 1%. The main causes of test rejection were haemolysis in case of serum samples, and clot formation and citrate excess in anticoagulated samples. The pneumatic tube transport resulted in significantly faster sample transport, more equalized sample arrival and processing, hence the turnaround time became shorter both for routine and emergency tests.
CONCLUSIONS: Autovalidation and proper control of preanalytical errors are essential for rapid and reliable laboratory service supported by the pneumatic tube system for sample transport.

Entities:  

Keywords:  hemolysis; hemolízis; laboratory diagnostics; laboratóriumi diagnosztika; leletátfordulási idő; pneumatic tube system; pneumatikus csőpostarendszer; preanalitikai hibák; preanalytical errors; turnaround time

Mesh:

Year:  2014        PMID: 25002315     DOI: 10.1556/OH.2014.29895

Source DB:  PubMed          Journal:  Orv Hetil        ISSN: 0030-6002            Impact factor:   0.540


  3 in total

Review 1.  Current Methods of Haemolysis Detection and Reporting as a Source of Risk to Patient Safety: a Narrative Review.

Authors:  Euan J McCaughey; Elia Vecellio; Rebecca Lake; Ling Li; Leslie Burnett; Douglas Chesher; Stephen Braye; Mark Mackay; Stephanie Gay; Tony C Badrick; Johanna I Westbrook; Andrew Georgiou
Journal:  Clin Biochem Rev       Date:  2016-12

2.  Clinical Laboratories - Production Factories or Specialized Diagnostic Centers.

Authors:  János Kappelmayer; Judit Tóth
Journal:  EJIFCC       Date:  2016-04-20

3.  Therapeutic turnaround times for common laboratory tests in a tertiary hospital in Kenya.

Authors:  Thomas Mwogi; Tim Mercer; Dan N Tina Tran; Ronald Tonui; Thorkild Tylleskar; Martin C Were
Journal:  PLoS One       Date:  2020-04-08       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.