Randomised masked clinical trial of the MGDRx EyeBag for the treatment of meibomian gland dysfunction-related evaporative dry eye.

Paramdeep Singh Bilkhu1, Shehzad Anjam Naroo1, James Stuart Wolffsohn1.   

Abstract

BACKGROUND/AIMS: To investigate the efficacy and safety of the MGDRx EyeBag (The Eyebag Company, Halifax, UK) eyelid warming device.
METHODS: Twenty-five patients with confirmed meibomian gland dysfunction (MGD)-related evaporative dry eye were enrolled into a randomised, single masked, contralateral clinical trial. Test eyes received a heated device; control eyes a non-heated device for 5 min twice a day for 2 weeks. Efficacy (ocular symptomology, non-invasive break-up time, lipid layer thickness, osmolarity, meibomian gland dropout and function) and safety (visual acuity, corneal topography, conjunctival hyperaemia and staining) measurements were taken at baseline and follow-up. Subsequent patient device usage and ocular comfort was ascertained at 6 months.
RESULTS: Differences between test and control eyes at baseline were not statistically significant for all measurements (p>0.05). After 2 weeks, statistically significant improvements occurred in all efficacy measurements in test eyes (p<0.05). Visual acuity and corneal topography were unaffected (p>0.05). All patients maintained higher ocular comfort after 6 months (p<0.05), although the benefit was greater in those who continued usage 1-8 times a month (p<0.001).
CONCLUSIONS: The MGDRx EyeBag is a safe and effective device for the treatment of MGD-related evaporative dry eye. Subjective benefit lasts at least 6 months, aided by occasional retreatment. TRIAL REGISTRATION NUMBER: NCT01870180. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.

Entities:  

Keywords:  Ocular surface; Tears; Treatment other

Mesh:

Year:  2014        PMID: 24997178      PMCID: PMC4251196          DOI: 10.1136/bjophthalmol-2014-305220

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


Introduction

Eyelid warming therapy is considered the mainstay of treating meibomian gland dysfunction (MGD).1 A variety of devices have been developed and shown to be efficacious in the treatment of MGD, but many are not commercially available.2–6 More recent studies focus on in-office treatment procedures, and have found long-term improvements in MGD,7–10 but these may be reserved for severe cases or those unresponsive to conventional procedures. There are no agreed-upon or standardised eyelid warming procedures in the scientific literature. Patient compliance issues relate to insufficient heating, duration and frequency of use.1 Studies which have attempted to optimise treatment using warm moist compresses suggest that precise and intense treatment is required to maintain sufficient eyelid warming,11 which would require high compliance and ability. However, the MGDRx EyeBag (eyebag), a commercially available eyelid warming device, may allow for improved compliance owing to its simple design. A recent study found that it produced significant increases in tear film lipid layer thickness and break-up time in healthy subjects.12 Increases in temperature sufficient to melt meibum1 was also observed on the upper and lower internal and external eyelids, suggesting it may be effective in treating MGD and evaporative dry eye caused by an ineffective tear film lipid layer.12 The aim of this study was to therefore investigate the efficacy and safety of the eyebag in patients with MGD-related evaporative dry eye.

Materials and methods

The study was a randomised, contralateral (one-test eye, fellow eye control), examiner-masked clinical trial, received ethical approval from the Aston University Research Ethics Committee and was registered as a clinical trial (NCT01870180: ClinicalTrials.gov). The research was conducted in accordance with the principles expressed in the Declaration of Helsinki.

Subjects

Subjects were identified during routine eye examination in a local primary care optometry practice, and were required to be ≥18 years old, with no systemic disease or medications known to affect the eyes, and no other active eye disease except symptomatic MGD-related evaporative dry eye. Diagnosis of at least mild severity was based upon recommendation by the diagnostic subcommittee of the International Workshop on Meibomian Gland Dysfunction: Ocular Surface Disease Index (OSDI) score of >12; presence of meibomian gland orifice plugging/obstruction on the lower or upper eyelids of both eyes; abnormal meibomian gland function in at least one eye (≤20 years old: quality or expressibility score >1; >20 years old: quality and expressibility score ≥1; see below).13 Non-invasive tear film break-up time (NITBUT) <10 s in at least one eye, and a negative Schirmer-1 test (>5.5 mm after 5 min) in at least one eye to differentiate between evaporative and aqueous deficiency dry eye.13 Eligible subjects were enrolled with written informed consent and attended for a baseline visit (Day 0) where the following battery of tests were performed in sequence.

Effectiveness measures

Ocular symptomology: dry eye symptoms were assessed for each eye using the OSDI. Meibomian gland function: based upon meibum quality and expressibility using a slit-lamp.13 Quality and expressibility of meibum from the central 8 meibomian glands on the lower and upper eyelid was graded on a 4-point scale (quality: 0=clear fluid, 1=cloudy fluid, 2=cloudy particulate fluid, 3=inspissated like toothpaste; expressibility: 0=all glands expressible, 1=3–4 glands expressible, 2=1–2 glands expressible, 3=no glands expressible) following firm digital pressure to the eyelid margins. Scores from the upper and lower eyelids were added to give a composite value for quality and expressibility. Meibomian gland dropout: measured via infrared meibography using the Keratograph 5M (Oculus Optikgeraete GmbH, Wetzlar, Germany) for the central 15 meibomian glands in the upper and lower everted eyelids of each eye. Meibomian gland dropout was graded using a 4-point scale (1=no partial glands; 2≤25% partial glands; 3=25–75% partial glands; 4≥75% partial glands). A partial gland was defined as one that is incomplete (presumed relative to the length of neighbouring intact meibomian glands) and present in lumps or clusters.13 Scores from the upper and lower eyelids were added to give a composite value. Tear film osmolarity: measured in each eye using the OcuSense TearLab (TearLab Corporation, San Diego, USA). Two measurements were taken in random order from each eye and were averaged to give a mean value for each eye.14 Tear film lipid layer thickness: measured non-invasively using a slit-lamp-mounted TearScope (Keeler, Windsor, UK) for each eye in random order. The thickness range of the tear film lipid layer was deduced from interference patterns observed using a standardised grading scale, where Grade 1=13–50 nm; Grade 2=30–50 nm; Grade 3=50–70 nm; Grade 4=80–90 nm; Grade 5=90–180 nm; Grade 6≥200 nm.15 Tear film stability: was determined by assessing the NITBUT using a slit-lamp-mounted TearScope on each eye in random order. NITBUT was defined as the time between the last complete blink and the appearance of a break or distortion in the fine grid pattern. This was repeated two more times, and the values were averaged to give a mean value.15 Tear Film Meniscus Height (TMH): measured using the Keratograph 5M. Images of the right and left eyes in primary gaze were captured immediately postblink. Digital callipers were used to measure the TMH located directly below the centre of the pupil, and was defined as the distance between the lower eyelid margin and the upper limit of the reflective zone—repeatability 95% CI±0.06 mm.

Safety measures

Visual acuity: best corrected monocular and binocular visual acuity was measured using a digital logMAR chart at 6 m, randomised between presentations (Test Chart 2000; Thomson Software Solutions, London, UK). Visual acuity was determined using letter by letter scoring with each letter corresponding to −0.02 logMAR units. Corneal topography: was measured on each eye using the Keratograph 5M. Conjunctival hyperaemia: digital images of the nasal and temporal bulbar conjunctiva were captured and analysed to provide an objective measurement via edge detection and colour extraction.16 Ocular surface staining: damage to the conjunctiva and cornea was assessed via instillation of lissamine green (GreenGlo, Sigma Pharmaceuticals, Monticello, USA) and fluorescein sodium (Fluorets, Chauvin Pharmaceuticals, London, UK) on each eye. Corneal, nasal and temporal bulbar conjunctival staining was graded individually using the Oxford system on a 6-point scale (0–5) to provide a composite score (0–15) for each eye.17

Intervention

Subjects were then randomised to receive a heated eyebag (40 s in microwave on full power) on either the left or right eye (test), chosen by random number generation. The contralateral eye received a non-heated eyebag (control). Subjects were instructed to use the eyebags at the same time twice a day (morning and evening, separated by at least 12 h) for 2 weeks (Days 1–14) and were required to provide feedback—subjects were texted twice a day (morning and evening) to grade ocular comfort for each eye on a 1–10 scale immediately before applying the eyebags (1=poor, 10=excellent). This score was also obtained at baseline. The examiner was masked to which eye received the heat treatment. Subjects then attended for follow-up (Day 15) where the baseline measurements were repeated, and were left with the eyebags to maintain their treatment if they felt it beneficial. Subsequent patient usage and reported ocular comfort (1–10 scale) was ascertained at 6±1 months.

Statistics

Statistical analysis was performed using SPSS for Microsoft Windows (Chicago, Illinois, USA). Differences between control and test eyes for visual acuity, ocular symptomology, corneal topography, conjunctival hyperaemia, and tear film stability, meniscus height, and osmolarity were evaluated by paired t tests, where the normal distribution was confirmed (Kolmogorov–Smirnov test: p>0.05). Differences between control and test eyes for non-normally distributed measures of tear film lipid layer thickness, meibomian gland function and dropout, and staining were evaluated by the Wilcoxon signed-rank test. Changes in ocular comfort scores over time were evaluated using Friedman's test, and posthoc analysis was performed using Wilcoxon signed-rank tests. To detect a primary treatment effect of 1 unit change in meibomian gland dropout (β=0.2, α=0.05), 23 subjects were required as each subject acted as their own control based upon published data (SD in MGD patients=±1.7 per eight glands).18 To account for possible drop-out, 25 patients (19 female, 6 male) were enrolled and completed the clinical trial (mean age: 28.7±7.8 years; range 19–42 years).

Results

At baseline, there was no statistically significant difference in ocular comfort scores between control and test eyes (mean score control eye: 5.15±1.57; test eye 5.25±1.33; Z=−0.30; p=0.76). An improvement in ocular comfort scores over time was observed over the 2 week treatment period in test eyes (X2=340.88, p<0.001) but not control eyes (χ2=27.78, p=0.42), with peak mean ocular comfort score in the test eye on the morning of Day 14 (8.55±0.89). A statistically significant difference in comfort scores between control and test eyes occurred from the evening on Day 1 (control eye: 5.00±1.45; test eye: 6.40±1.35; Z=−3.72, p<0.001), and this difference remained statistically significant (p<0.001) for the entire treatment period (figure 1).
Figure 1

Mean ocular comfort scores for control eyes and test eyes during the treatment period (Day 1 to Day 14). Error bars represent 1 SD. The difference in ocular comfort scores between control and test eyes was statistically significant (p<0.001) from the evening of Day 1 onwards.

Mean ocular comfort scores for control eyes and test eyes during the treatment period (Day 1 to Day 14). Error bars represent 1 SD. The difference in ocular comfort scores between control and test eyes was statistically significant (p<0.001) from the evening of Day 1 onwards. The difference between test and control eyes at baseline was not statistically significant for all measurements (p>0.05; table 1). After the treatment period, there was a statistically significant improvement in all efficacy measurements in test eyes only (p<0.05), except for meibomian gland expressibility in control eyes (p<0.05). Visual acuity and corneal topography were unaffected (p>0.05), whereas conjunctival hyperaemia and ocular surface staining decreased in test eyes (table 1).
Table 1

Mean (±1 SD) values for all efficacy and safety parameters for test and control eyes at baseline and follow-up

 BaselineFollow-upTest eye compared to baseline (p)
ControlTestSignificanceControlTestSignificance
OSDI43.0±14.443.9±13.40.3639.8±12.720.7±8.7<0.001*<0.001*
Meibomian gland quality2.40±1.092.35±0.990.762.05±1.231.3±0.92<0.05*<0.001*
Meibomian gland expressibilty2.00±1.082.15±0.890.261.70±1.031.35±0.82<0.05*<0.001*
Meibomian gland drop-out2.80±1.242.90±1.250.412.65±1.042.05±1.15<0.05*<0.05*
Osmolarity (mOsm/L)310.3±10.6310.9±9.50.43308.9±11.6305.0±9.1<0.05*<0.001*
Lipid layer thickness2.10±0.722.05±0.890.662.2±0.623.1±0.91<0.05*<0.001*
Tear film stability (s)9.29±3.649.48±3.740.129.32±3.5611.37±3.45<0.001*<0.001*
Tear film meniscus height (mm)0.34±0.060.34±0.070.690.34±0.060.36±0.06<0.001*<0.05*
Visual acuity (LogMAR)−0.04±0.03−0.04±0.030.79−0.05±0.03−0.05±0.030.420.06
Corneal topography (e)0.37±0.150.37±0.160.890.37±0.150.37±0.170.910.65
Conjunctival hyperaemia vessel coverage1.86±1.521.61±1.190.281.44±1.060.94±0.65<0.05*<0.05*
Conjunctival hyperaemia coloration38.81±1.6238.53±1.660.2338.49±1.4238.00±1.41<0.05*<0.05*
Ocular surface staining3.2±2.063.75±1.710.072.95±1.391.85±0.99<0.05*<0.001*

Asterisks (*) represent statistical significance. Note: conjunctival vessel coverage and colouration an average of superior, inferior, nasal and temporal quadrants as the effect was similar across ocular surface regions.

Mean (±1 SD) values for all efficacy and safety parameters for test and control eyes at baseline and follow-up Asterisks (*) represent statistical significance. Note: conjunctival vessel coverage and colouration an average of superior, inferior, nasal and temporal quadrants as the effect was similar across ocular surface regions. After 6 months, four patients reported having stopped using the eyebag, but their ocular comfort was still better than baseline (6.20±1.03 vs 4.80±1.40, p=0.039). Patients (n=8) who continued to use the treatment for 1–4 weeks had a greater ocular comfort benefit (7.79±0.89 vs 5.64±1.33, p<0.001), but the greatest scores were achieved by those who continued to use the eyebag 1–8 times a month (7.96±1.46 vs 5.12±1.50, p<0.001). One patient reported a transient stinging sensation when the heated eyebag was placed on the upper eyelid on the first four occasions. No other adverse event was reported by this patient or the others for the duration of treatment period.

Discussion

The results demonstrate statistically significant improvements in meibomian gland dropout, TMH, osmolarity, conjunctival hyperaemia and staining in test eyes only (table 1). Given that measurements were taken on two separate occasions, it was not possible to determine a natural history of the treatment approach. It is likely that the frequent and regular heating melted the abnormal meibum, clearing the obstruction within the meibomian gland, allowing the meibum to secrete on the eyelid margin.19 The increased quantity of meibum available may, therefore, help thicken and restore normal tear film lipid layer function such that evaporation of the underlying aqueous is prevented, tear film stability is improved, tear film osmolarity and ocular surface staining reduced and tear film meniscus height increased.20 This is supported by the significant positive correlation observed between lipid layer thickness and tear film stability.21 Furthermore, meibum quantity or evaporation rates have previously been shown to positively correlate with tear film stability.22 Indeed, NITBUT (by 1.9±1.3 s) and tear film lipid layer thickness (by 1.1±0.8 grades) demonstrated a statistically significant increase following heat treatment. Although a recent study investigating the effect of warm moist air goggles in MGD patients found no statistically significant increase in tear film stability 10 min after use,23 Olson et al (2003) found an 80% increase in lipid layer thickness in MGD-related dry eye patients after 5 min into treatment with warm moist compress heated to 40°C for 30 min, and 66% increase 5 min after removal.3 Thus, it may be that improvements in tear film stability and lipid layer thickness in MGD patients can only be achieved where eyelid warming devices make contact with the eyelids. Meibomian gland function also improved with the heated eyebag after the treatment period, where there were statistically significant improvements in meibum quality and expressibility in test eyes. However, the improvement in meibum quality compared to control eyes was only statistically significant between the lower eyelids. At baseline, meibum quality grade on the lower eyelid was significantly greater than the upper eyelid of test and control eyes, such that the upper eyelid was considered normal—thus, improvements in meibum quality may only have been detected in the lower eyelids owing to the limited resolution of the grading scale. Of interest was the statistically significant improvement in expressibility in control eyes (composite reduction of 0.35±0.81). The non-heated eyebag may have warmed the eyelids by insulating and preventing heat loss, which subsequently brought about an improvement in the abnormal meibum. The frequent application of the eyebag with the hands may have also massaged the eyelids to help clear the obstructed glands. Since the heated and non-heated eyebags were applied at the same time and duration, this massaging effect may have contributed at least in part to the greater improvement in expressibility observed in test eyes (composite reduction of 0.80±0.70). With respect to the semiquantitative nature of the grading scale, eyelids with 5, 6, or 7 expressible glands were recorded as ‘grade 1’—thus, the improvements observed may have been underestimated. Furthermore, a statistically significant improvement in ocular symptoms was also observed in test eyes only. Surprisingly, the improvement in ocular comfort scores was observed as early as the evening of treatment Day 1, suggesting that the eyebag produces an increase in ocular comfort for at least 12 h following single use. Although subjects still had raised ocular comfort at around 6 months after the trial, comfort was maintained best in those who continued treatment 1–8 times a month. It should be noted that subjects were not masked and, therefore, some placebo effect could be possible, although the effect was marked and occurred in all subjects. Eyelid warming therapy associated with eyelid massage has been suggested to induce corneal deformation due to increases in corneal temperature, and with concurrent massaging, possible ectasia.24 Only one case of corneal deformation following such treatment has been reported, but the duration of treatment in this case was far longer (15 min twice a day for 7 weeks) than that prescribed herein. Transient visual degradation without changes in corneal topography has been observed after warm compresses (45°C) were applied every 2 min for 30 min without massage,25 but again the duration of treatment and peak temperature was sustained for longer than the present study and the treatment method is not typically advised to MGD patients. Therefore, it appears that corneal deformation and/or visual changes, may only occur following unusually long and intense treatment application durations and long-term therapy associated with eyelid rubbing. Given there was no reduction in visual acuity and no change in corneal topography from baseline after the treatment period, the risk of corneal deformation with the MGDRx EyeBag, as prescribed, appears to be minimal. Further, there was a statistically significant improvement in ocular surface staining (particularly the cornea) and conjunctival hyperaemia in test eyes. Thus, when used based upon the treatment regimen prescribed herein, the MGDRx EyeBag may be considered a safe and effective eyelid warming device with a low risk of corneal deformation and visual changes, and results in improved comfort and tear film parameters involved in evaporative dry eye in patients with MGD.
  25 in total

1.  A novel thermodynamic treatment for meibomian gland dysfunction.

Authors:  Beth R Friedland; Christopher P Fleming; Caroline A Blackie; Donald R Korb
Journal:  Curr Eye Res       Date:  2011-02       Impact factor: 2.424

2.  Inner eyelid surface temperature as a function of warm compress methodology.

Authors:  Caroline A Blackie; Jessica D Solomon; Jack V Greiner; Mycroft Holmes; Donald R Korb
Journal:  Optom Vis Sci       Date:  2008-08       Impact factor: 1.973

3.  Tear film with "Orgahexa EyeMasks" in patients with meibomian gland dysfunction.

Authors:  Reiko Ishida; Yukihiro Matsumoto; Tatsuya Onguchi; Minako Kaido; Kenji Iwamuro; Julie Kobayashi; Yoji Takano; Jun Shimazaki; Eiki Goto; Murat Dogru; Kazuo Tsubota
Journal:  Optom Vis Sci       Date:  2008-08       Impact factor: 1.973

Review 4.  The international workshop on meibomian gland dysfunction: report of the subcommittee on management and treatment of meibomian gland dysfunction.

Authors:  Gerd Geerling; Joseph Tauber; Christophe Baudouin; Eiki Goto; Yukihiro Matsumoto; Terrence O'Brien; Maurizio Rolando; Kazuo Tsubota; Kelly K Nichols
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-03-30       Impact factor: 4.799

5.  Comparison of human tear film osmolarity measured by electrical impedance and freezing point depression techniques.

Authors:  Alan Tomlinson; Louise C McCann; Edward I Pearce
Journal:  Cornea       Date:  2010-09       Impact factor: 2.651

6.  Long-term (12-month) improvement in meibomian gland function and reduced dry eye symptoms with a single thermal pulsation treatment.

Authors:  Jack V Greiner
Journal:  Clin Exp Ophthalmol       Date:  2012-12-14       Impact factor: 4.207

7.  A new system, the LipiFlow, for the treatment of meibomian gland dysfunction.

Authors:  Stephen S Lane; Harvey B DuBiner; Randy J Epstein; Paul H Ernest; Jack V Greiner; David R Hardten; Edward J Holland; Michael A Lemp; James E McDonald; David I Silbert; Caroline A Blackie; Christy A Stevens; Raman Bedi
Journal:  Cornea       Date:  2012-04       Impact factor: 2.651

8.  A single LipiFlow® Thermal Pulsation System treatment improves meibomian gland function and reduces dry eye symptoms for 9 months.

Authors:  Jack V Greiner
Journal:  Curr Eye Res       Date:  2012-02-10       Impact factor: 2.424

9.  Meibomian gland dysfunction in chronic blepharitis.

Authors:  W D Mathers; W J Shields; M S Sachdev; W M Petroll; J V Jester
Journal:  Cornea       Date:  1991-07       Impact factor: 2.651

10.  Disposable eyelid-warming device for the treatment of meibomian gland dysfunction.

Authors:  Asako Mori; Jun Shimazaki; Shigeto Shimmura; Hiroshi Fujishima; Yoshihisa Oguchi; Kazuo Tsubota
Journal:  Jpn J Ophthalmol       Date:  2003 Nov-Dec       Impact factor: 2.447

View more
  10 in total

Review 1.  Revisiting the vicious circle of dry eye disease: a focus on the pathophysiology of meibomian gland dysfunction.

Authors:  Christophe Baudouin; Elisabeth M Messmer; Pasquale Aragona; Gerd Geerling; Yonca A Akova; José Benítez-del-Castillo; Kostas G Boboridis; Jesús Merayo-Lloves; Maurizio Rolando; Marc Labetoulle
Journal:  Br J Ophthalmol       Date:  2016-01-18       Impact factor: 4.638

2.  Impact of Chronic Smoking on Meibomian Gland Dysfunction.

Authors:  Shen Wang; Hui Zhao; Caihong Huang; Zhengri Li; Wei Li; Xiaobo Zhang; Zuguo Liu
Journal:  PLoS One       Date:  2016-12-28       Impact factor: 3.240

3.  Therapies for Meibomian Gland Dysfunction: A Systematic Review and Meta-Analysis of Randomized Controlled Trials.

Authors:  Kannan Sridharan; Gowri Sivaramakrishnan
Journal:  Open Ophthalmol J       Date:  2017-11-22

4.  Effects of a warm compress containing menthol on the tear film in healthy subjects and dry eye patients.

Authors:  Reiko Arita; Naoyuki Morishige; Ichiro Sakamoto; Natsuko Imai; Yuko Shimada; Michihito Igaki; Atsushi Suzuki; Kouzo Itoh; Kazuo Tsubota
Journal:  Sci Rep       Date:  2017-04-05       Impact factor: 4.379

5.  TheraPearl Eye Mask and Blephasteam for the treatment of meibomian gland dysfunction: a randomized, comparative clinical trial.

Authors:  Jonatan Olafsson; Xiaoran Lai; Erlend Christoffer Sommer Landsend; Snorri Olafsson; Eric Parissi; Øygunn A Utheim; Sten Raeder; Reza A Badian; Neil Lagali; Darlene A Dartt; Tor P Utheim
Journal:  Sci Rep       Date:  2021-11-17       Impact factor: 4.379

Review 6.  Eyelid Warming Devices: Safety, Efficacy, and Place in Therapy.

Authors:  Brandon Bzovey; William Ngo
Journal:  Clin Optom (Auckl)       Date:  2022-08-05

Review 7.  A Review of Imaging Biomarkers of the Ocular Surface.

Authors:  William W Binotti; Betul Bayraktutar; M Cuneyt Ozmen; Stephanie M Cox; Pedram Hamrah
Journal:  Eye Contact Lens       Date:  2020-03       Impact factor: 3.152

8.  Short-term effect of a developed warming moist chamber goggle for video display terminal-associated dry eye.

Authors:  Yueping Ren; Jie Chen; Qinxiang Zheng; Wei Chen
Journal:  BMC Ophthalmol       Date:  2018-02-07       Impact factor: 2.209

9.  An Eyelid Warming Device for the Management of Meibomian Gland Dysfunction.

Authors:  William Ngo; Sruthi Srinivasan; Lyndon Jones
Journal:  J Optom       Date:  2018-10-16

Review 10.  Non-pharmaceutical treatment options for meibomian gland dysfunction.

Authors:  Reiko Arita; Shima Fukuoka
Journal:  Clin Exp Optom       Date:  2020-01-13       Impact factor: 2.742

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.