Literature DB >> 24997109

Rapid and simultaneous determination of sulfonate ester genotoxic impurities in drug substance by liquid chromatography coupled to tandem mass spectrometry: comparison of different ionization modes.

Tian Guo1, Yuanyuan Shi1, Li Zheng1, Feng Feng2, Feng Zheng3, Wenyuan Liu4.   

Abstract

Two ionization techniques for liquid chromatography-mass spectrometry (LC-MS) determination of sulfonate ester potentially genotoxic impurities (PGIs) were evaluated. Twelve PGIs including methyl, ethyl, propyl and isopropyl esters of methanesulfonate, benzenesulfonate and p-toluenesulfonate were studied in this research. Electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) sources were compared in terms of performance and quality parameters for detection of the twelve PGIs. Their mass spectra obtained by APCI and ESI were very different in both fragment ions and relative abundances. In APCI negative ion mode the twelve sulfonate esters showed their stable precursor ions of [M-alkyl](-), which readily yielded product ions of [M-alkyl-CH3](-) (for aliphatic sulfonate esters) or [M-alkyl-SO2](-) (for aromatic sulfonate esters) with collision-induced dissociation (CID) applied; and working in selected reaction monitoring (SRM) mode has allowed limits of detection to be decreased. In the case of ESI ionization, these compounds showed their precursor ions [M+H](+), but their abundance was easily competed by formation of ions [M+NH4](+) and/or [M+Na](+), which led to poor analytical sensitivity and reproducibility. Although mobile phase additives could enhance the responses of adduct ions like [M+NH4](+) and [M+Na](+), no improvement was obtained when using SRM mode. Twelve sulfonate esters were systematically compared and APCI was shown to be a better ionization technique for rapid and sensitive determination of these PGIs. Performance of the developed approach for rapid determination of 12 PGIs was also evaluated. Quality parameters were established and good precision (relative standard deviations <8%) and very low limits (2-4ng/mL) of detection were obtained, mainly when using APCI in SRM mode.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Atmospheric pressure chemical ionization; Comparison of ionization modes; Electrospray ionization; Rapid and simultaneous determination; Sulfonate esters

Mesh:

Substances:

Year:  2014        PMID: 24997109     DOI: 10.1016/j.chroma.2014.05.079

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  4 in total

1.  Quantitative Determination of Four Potential Genotoxic Impurities in the Active Pharmaceutical Ingredients in TSD-1 Using UPLC-MS/MS.

Authors:  Taiyu Wang; Hailong Yang; Jie Yang; Ningjie Guo; Guodong Wu; Xueyu Xu; Ming An
Journal:  Molecules       Date:  2022-06-27       Impact factor: 4.927

2.  Determination of Residual Solvents in Linezolid by Static Headspace GC.

Authors:  Xiao-Zhen Feng; Guo-Cheng Han; Jianhe Qin; Shi-Min Yin; Zhencheng Chen
Journal:  J Chromatogr Sci       Date:  2015-12-10       Impact factor: 1.618

3.  Simultaneous Determination of 15 Sulfonate Ester Impurities in Phentolamine Mesylate, Amlodipine Besylate, and Tosufloxacin Tosylate by LC-APCI-MS/MS.

Authors:  Bo Jin; Kaijing Guo; Tingting Zhang; Tong Li; Chen Ma
Journal:  J Anal Methods Chem       Date:  2019-10-07       Impact factor: 2.193

4.  Determination of Methyl Methanesulfonate and Ethyl Methylsulfonate in New Drug for the Treatment of Fatty Liver Using Derivatization Followed by High-Performance Liquid Chromatography with Ultraviolet Detection.

Authors:  Yue Wang; Jing Feng; Song Wu; Huihui Shao; Wenxuan Zhang; Kun Zhang; Hanyilan Zhang; Qingyun Yang
Journal:  Molecules       Date:  2022-03-17       Impact factor: 4.411

  4 in total