D J Corvan1, G Sarri1, M Zepf1.
Abstract
A novel design for a compact gamma-ray spectrometer is presented. The proposed system allows for spectroscopy of high-flux multi-MeV gamma-ray beams with MeV energy resolution in a compact design. In its basic configuration, the spectrometer exploits conversion of gamma-rays into electrons via Compton scattering in a low-Z material. The scattered electron population is then spectrally resolved using a magnetic spectrometer. The detector is shown to be effective for gamma-ray energies between 3 and 20 MeV. The main properties of the spectrometer are confirmed by Monte Carlo simulations.Entities:
Year: 2014 PMID: 24985864 DOI: 10.1063/1.4884643
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523