Modifications to a commercially available linear mass spectrometer for mass-resolved microscopy with the pixel imaging mass spectrometry (PImMS) camera.

E Halford1, B Winter, M D Mills, S P Thompson, V Parr, J J John, A Nomerotski, C Vallance, R Turchetta, M Brouard.   

Abstract

RATIONALE: Imaging mass spectrometry is a powerful analytical technique capable of accessing a large volume of spatially resolved, chemical data from two-dimensional samples. Probing the entire surface of a sample simultaneously requires a detector with high spatial and temporal resolutions, and the ability to observe events relating to different mass-to-charge ratios.
METHODS: A commercially available time-of-flight mass spectrometer, designed for matrix-assisted laser desorption/ionization (MALDI) analysis, was combined with the novel pixel imaging mass spectrometry (PImMS) camera in order to perform multi-mass, microscope-mode imaging experiments. A number of minor modifications were made to the spectrometer hardware and ion optics so that spatial imaging was achieved for a number of small molecules.
RESULTS: It was shown that a peak width of Δm50 %  <  1  m/z unit across the range 200 ≤ m/z  ≤  800 can be obtained while also achieving an optimum spatial resolution of 25 µm. It was further shown that these data were obtained simultaneously for all analytes present without the need to scan the experimental parameters.
CONCLUSIONS: This work demonstrates the capability of multi-mass, microscope-mode imaging to reduce the acquisition time of spatially distributed analytes such as multi-arrays or biological tissue sections. It also shows that such an instrument can be commissioned by effecting relatively minor modifications to a conventional commercial machine.
Copyright © 2014 John Wiley & Sons, Ltd.

Entities:  

Mesh:

Year:  2014        PMID: 24975244     DOI: 10.1002/rcm.6940

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  1 in total

1.  MALDI MS Imaging at Acquisition Rates Exceeding 100 Pixels per Second.

Authors:  Antonín Bednařík; Markéta Machálková; Eugene Moskovets; Kateřina Coufalíková; Pavel Krásenský; Pavel Houška; Jiří Kroupa; Jarmila Navrátilová; Jan Šmarda; Jan Preisler
Journal:  J Am Soc Mass Spectrom       Date:  2018-11-19       Impact factor: 3.109

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.